Abstract
Objectives Continuous hemodynamic monitoring offers the opportunity to individualize management in severe preeclampsia (PEC). We compared cardiac output (CO) and systemic vascular resistance (SVR) measured by bioreactance (NICOM), Clearsite™ Fingercuff [CS), and 3D-echocardiography (3DE).
Study Design This prospective observational study included 12 pregnant patients with early PEC. CO and SVR were measured simultaneously by NICOM, CS, and 3DE antepartum and 1-2 days postpartum. Using 3DE as the standard, CS and NICOM interchangeability, precision, accuracy, and correlation were assessed.
Results Compared to 3DE-CO, CS-CO was highly correlated (R2=0.70, p=<0.0001) with low percentage error (PE 29%) which met criteria for interchangeablity. CS-SVR had strong correlation (R2=0.81, p=<0.0001) and low PE (29%). While CS tended to slightly overestimate CO (bias +2.05 +/-1.18 L/min, limit of agreement (LOA) -0.20-4.31) and underestimate SVR (bias -279 +/-156 dyes/sec/cm5; LOA -580-18.4) these differences were unlikely to be clinically significant. Thus CS could be interchangeable with 3DE for CO and SVR. NICOM-CO had only moderate correlation with 3DE-CO (R2=0.29, p=0.01) with high PE (52%) above threshold for interchangeability. NICOM-CO had low mean bias (−1.2 +/-1.68 L/min) but wide 95% LOA (−4.41 - 2.14) suggesting adequate accuracy but low precision in relation to 3DE-CO. NICOM-SVR had poor correlation with 3DE-SVR (R2=0.32, p=0.001) with high PE (67%), relatively low mean bias (238 +/-256), and wide 95% LOA (– 655–1131). NICOM did not meet the criteria for interchangeable with 3DE for CO and SVR.
Conclusions Clearsite Fingercuff, but not NICOM, has potential to be clinically useful for CO and SVR monitoring in severe preeclampsia.
Competing Interest Statement
Dr Josephine Chou is a consultant for abiomed. The remaining authors declare no conflicts of interest.
Funding Statement
The research reported in this publication was supported by the Bahn endowment fund from Yale University School of Medicine and the Garite Mini-Sabbatical Grant from the Society for Maternal Fetal Medicine.
Author Declarations
I confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.
Yes
The details of the IRB/oversight body that provided approval or exemption for the research described are given below:
The Yale School of Medicine Human Investigation Committee ethical approval for this study. All patients signed an informed consent prior to enrollment.
I confirm that all necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived, and that any patient/participant/sample identifiers included were not known to anyone (e.g., hospital staff, patients or participants themselves) outside the research group so cannot be used to identify individuals.
Yes
I understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).
Yes
I have followed all appropriate research reporting guidelines, such as any relevant EQUATOR Network research reporting checklist(s) and other pertinent material, if applicable.
Yes
Footnotes
U246381{at}bcm.edu
jbhinder{at}northwell.edu, Lsugeng{at}northwell.edu
maxwell.eder{at}yale.edu
paul.heerdt{at}yale.edu
hlipkind{at}med.cornell.edu
Ur2137{at}cumc.columbia.edu
Data Availability
All data produced in the present study are available upon reasonable request to the authors.
ABBREVIATIONS:1
- 1PEC
- severe preeclampsia
- CO
- cardiac output
- SVR
- systemic vascular resistance
- CS
- Clearsite Fingercuff
- NICOM
- non-invasive cardiac monitoring with bioreactance
- 3DE3
- 3D-echocardiography
- PE
- percentage error
- LOA
- limit of agreement
- HDP
- hypertensive disorders of pregnancy
- SBP
- systolic blood pressure
- DBP
- diastolic blood pressure
- MAP
- mean arterial pressures
- SV
- stroke volume
- CVP
- central venous pressure
- LV
- left ventricular
- EF
- ejection fraction
- BP
- blood pressure
- GLS
- global longitudinal strain