Abstract
By the first week of April 2020, more than 1,500,000 positive cases of COVID-19 and more than 50,000 deaths had been officially reported worldwide. While developed countries such as the USA, Italy, England, France, Spain, and Germany struggle to mitigate the propagation of SARS-CoV-2, the COVID-19 pandemic arrived in Latin America, India, and Africa—territories in which the mounted infrastructure for diagnosis is greatly underdeveloped. An actual epidemic emergency does not provide the required timeframe for testing new diagnostic strategies; therefore, the first line of response must be based on commercially and readily available resources. Here, we demonstrate the combined use of a three-dimensional (3D)-printed incubation chamber for commercial Eppendorf PCR tubes, and a colorimetric embodiment of a loop-mediated isothermal amplification (LAMP) reaction scheme for the detection of SARS-CoV-2 nucleic acids. We used this strategy to detect and amplify SARS-CoV-2 DNA sequences using a set of in-house designed initiators that target regions encoding the N protein. We were able to detect and amplify SARS-CoV-2 nucleic acids in the range of ∼625 to 2 × 105 DNA copies by this straightforward method. The accuracy and simplicity of this diagnostics strategy may provide a cost-efficient and reliable alternative for use during the COVID-19 pandemics, particularly in underdeveloped regions were the availability of RT-qPCR instruments may be limited. Moreover, the portability, ease of use, and reproducibility of this strategy make it a reliable alternative for deployment of point-of-care SARS-CoV-2 detection efforts during the pandemics.
Introduction
By the end of the first week of April 2020, more than one and a half million positive cases of COVID-19 were officially reported across the globe[1]. Even developed countries, such as the USA, England, France, and Germany, are struggling to mitigate the propagation of SARS-CoV-2 by implementing social distancing and widespread testing. Less developed regions, such as Latin America, India, and Africa, are now experiencing the arrival of COVID-19, but these—territories are woefully lacking in the finances or the mounted infrastructure for diagnosis of this pandemic infection. Rapid and massive testing of thousands of possibly infected subjects has been an important component of the strategy of the countries that are effectively mitigating the spreading of COVID-19 among their populations (i.e., China[2], South Korea [3], and Singapore [4]). By comparison, developing countries with high demographic densities, such as México [5], India [6], or Brazil [7], may not be able to implement a sufficient number of centralized laboratories for rapid large-scale testing for COVID-19.
Many methodologies have been proposed recently to deliver cost-effective diagnosis (i.e., those based on immunoassays [8–11] or specific gene hybridization assisted by CRISPR-Cas systems [12–14]). While immunoassays are an accurate and efficacious tool for assessing the extent of the infection for epidemiological studies [15], their usefulness is limited to the identification of infected subjects during early phases of infection [11,16], a critical period for infectiveness. For instance, experimental evidence collected from a small number of COVID-19 patients (9 subjects) showed that 100% of them produced specific immunoglobulins G (IgGs) for SARS-CoV-2 within two weeks of infection, but only 50% of them did during the first week post infection [17].
Nucleic acid amplification continues to be the gold standard for the detection of viral diseases in the early stages [18–22], and very small viral loads present in symptomatic or asymptomatic patients can be reliably detected using amplification based technics, such as PCR[23–25], RPA[26], and LAMP[27–29].
During the last two pandemic events with influenza A/H1N1/2009 and COVID-19, the Centers for Disease Control (CDC) and the World Health Organization (WHO) recommended real-time quantitative PCR (RT-qPCR) methods as the gold standard for official detection of positive cases[16,30]. However, the reliance on RT-qPCR often leads to dependence on centralized laboratory facilities for testing [16,30–33]. To resolve this drawback, isothermal amplification reaction schemes (i.e., loop-mediated isothermal amplification (LAMP) and recombinase polymerase amplification (RPA)) have been proposed as alternatives to PCR-based methods and devices for point-of-care settings [32,34,35]. The urgency of using reliable molecular-based POC methods for massive diagnostic during epidemiological emergencies has become even more evident during the current COVID-19 pandemics [30,36,37].
In these times of COVID-19 [38], scientists and philanthropists around the globe have worked expeditiously on the development of rapid and portable diagnostics for SARS-CoV-2. Several reports have demonstrated the use of colorimetric LAMP-based methods for diagnosis of pandemic COVID-19 [39–44]. Some of these reports (currently available as preprints) use phenol red, a well-known pH indicator, to assist in the visual discrimination between positive and negative samples [39,40].
In this study, we demonstrate the use of a simple embodiment of a colorimetric Loop-mediated Isothermal amplification (LAMP) protocol for the detection and amplification of synthetic samples of SARS-CoV-2, the causal viral agent of COVID-D. In this LAMP-based strategy, also assisted by the use of phenol red, sample incubation is greatly facilitated by the use of a three-dimensional (3D)-printed incubator connected to a conventional water circulator, while discrimination between positive and negative samples is achieved by visual inspection. We quantitatively analyze differences in color between positive and negative samples using color decomposition and analysis in the color CIELab space[45]. Moreover, we compare the sensibility of this LAMP colorimetric method versus PCR protocols. This simple strategy is potentially adequate for the fast deployment of diagnostic efforts in the context of COVID-19 pandemics.
Materials and Methods
Equipment specifications
We ran several hundred amplification experiments using a colorimetric LAMP method in a 3D-printed incubation chamber designed in house and connected to a conventional water circulator (Figure 1). The design and all dimensional specifications of this chamber have been made available in Supplementary Information (Figure S1,S2; Supplementary File S1). In the experiments reported here, we used a chamber with dimensions of 20 × 5 × 15 cm3 and a weight of 0.4 kg (without water). A conventional water circulator (WVR, PA, USA), was used to circulate hot water (set point value at 76 °C) through the 3D-printed chamber for incubation of the Eppendorf PCR tubes (0.2 mL). In this first chamber prototype, twelve amplification reactions can be run in parallel. This concept design is amenable for fabrication in any STL-3D printing platform and may be scaled up to accommodate a larger number of tubes.
We used a blueGel electrophoresis unit, powered by 120 AC volts, to validate the LAMP amplification using gel electrophoresis. Photo-documentation was done using a smartphone camera. We also used a Synergy HT microplate reader (BioTek Instruments, VT, USA) to detect the fluorescence induced by an intercalating reagent in positive samples from the PCR reactions.
Validation DNA templates
We used plasmids containing the complete N gene from 2019-nCoV, SARS, and MERS as positive controls, with a concentration of 200,000 copies/µL (Integrated DNA Technologies, IA, USA).
Samples containing different concentrations of synthetic nucleic acids of SARS-CoV-2 were prepared by successive dilutions from stocks (from 2 × 105 copies to 65 copies). We used a plasmid that contained the gene GP from EBOV as a negative control. The production of this EBOV genetic material has been documented elsewhere by our group [23].
Amplification mix
We used WarmStart® Colorimetric LAMP 2× Master Mix (DNA & RNA) from New England Biolabs (MA, USA), and followed the recommended protocol: 12.5 μL Readymix, 1.6 μM FIP primer, 1.6 μM BIP primer, 0.2 μM F3 primer, 0.2 μM B3 primer, 0.4 μM LF primer, 0.4 μM LB primer, 1μL DNA template (∼ 625 to 2 × 105 DNA copies), 1.25 μL EvaGreen® Dye from Biotium (CA, USA), and nuclease-free water to a final volume of reaction 25 μL. This commercial mix contains phenol red as a pH indicator for revealing the shift of pH during LAMP amplification across the threshold of pH=6.8.
Primers used
Two different sets of LAMP primers, referred to here as α and β, were designed in house using the LAMP primer design software Primer Explorer V5 (http://primerexplorer.jp/lampv5e/index.html). These primers were based on the analysis of alignments of the SARS-Co2 N gene sequences using the software Geneious (New Zealand), downloaded from https://www.ncbi.nlm.nih.gov/genbank/sars-cov-2-seqs/#nucleotide-sequences.
Each set, containing six LAMP primers, were used to target two different regions of the sequence of the SARS-Co2 N gene. In addition, for comparison purposes, we conducted PCR amplification experiments using the primer sets recommended by the CDC for the standard diagnostics of COVID-19 (i.e., N1, N2, and N3 assays) using RT-qPCR. The sequences of our LAMP primers are presented in Table 1. The sequences of the PCR primers (N1–N3) have been reported elsewhere[24,46].
Amplification protocols
For all LAMP experiments, we performed isothermal heating for 30 or 60 min. In our experiments, we tested three different temperatures: 50, 60, and 65 °C.
Documentation of LAMP products
We analyzed 10 μL of each LAMP reaction in a blueGel unit, a portable electrophoresis unit sold by MiniPCR from Amplyus (MA, USA). This is a compact electrophoresis unit (23 × 10 × 7 cm) that weighs 350 g. In these experiments, we analyzed 10 μL of the LAMP product using 1.2 % agarose electrophoresis tris-borate-EDTA buffer (TBE). We used the Quick-Load Purple 2-Log DNA Ladder (NEB, MA, USA) as a molecular weight marker. Gels were dyed with Gel-Green from Biotium (CA, USA) using a 1:10,000 dilution, and a current of 48 V was supplied by the blueGel built-in power supply (AC 100–240V, 50–60Hz).
As an alternative method for detection and reading of the amplification product, we evaluated the amplification products by detecting the fluorescence emitted by a DNA-intercalating agent, the EvaGreen® Dye from Biotium (CA, USA), in a Synergy HT microplate reader (BioTek Instruments, VT, USA). Briefly, 25 μL of the LAMP reaction was placed in separate wells of a 96-well plate following completion of the LAMP incubation. A 125 μL volume of nuclease-free water was added to each well for a final sample volume of 150 μL and the samples were well-mixed by pipetting. These experiments were run in triplicate. The following conditions were used in the microplate reader: excitation of 485/20, emission of 528/20, gain of 75. Fluorescence readings were done from the top at room temperature.
Color determination by image analysis
We also photographically documented and analyzed the progression of color changes in the positive and negative SARS-CoV-2 synthetic samples during the LAMP reaction time (i.e., from 0 to 50 min). For that purpose, Eppendorf PCR tubes containing LAMP samples were photographed using a smartphone (iPhone, from Apple, USA). We used an application for IOS (Color Companion, freely available at Apple store) to determine the components of color of each LAMP sample in the CIELab color space. Color differences between the positive samples and negative controls were calculated as distances in the CIELab coordinate system according to the following formula:
Here L, a, and b are the color components of the sample or the negative control in the CIELab color space (Supplementary Figure 4).
Results and Discussion
Rationale
We have developed a simple diagnostic method for the detection of SARS-CoV-2, the causal agent of COVID-19. The rationale underlying this strategy is centered on achieving the simplest possible integration of easily available reagents, materials, and fabrication techniques to facilitate fast and massive implementation during the current COVID-19 pandemics in low- or middle-income regions.
This method is based on the amplification of the genetic material of SARS-CoV-2 using a loop-mediated isothermal amplification (LAMP). The amplification is conducted using a commercial reaction mix in commercial and widely available 200 µL Eppendorf PCR tubes. Moreover, we have designed and fabricated a simple 3D-printed chamber (Figure 1) for incubation of the Eppendorf tubes and to enable LAMP at high temperatures (50–65 °C) and extended times (up to 1 h). We show that this incubation chamber, when connected to a conventional water recirculator, enables the successful amplification of positive samples (i.e., samples containing SARS-CoV-2 nucleic acids).
This incubation chamber is one of the key elements that enable rapid and widespread implementation of this diagnostic method at low cost. This 3D-printed incubator can be rapidly printed using standard SLA printers widely available in markets worldwide.
Standard 3D-printing resins can be used. The availability of the original AutoCAD files (included here as supplemental material) enables fast modification/optimization of the design for accommodation of a larger number of samples or larger or smaller tubes, adaptation to any available hoses (tubing), and possible incorporation of an on-line color reading system. Indeed, all this is consistent with the main rationale of our proposed diagnostic strategy for pandemic COVID-19: To enabling a fast and feasible response using widespread, distributed, and scalable diagnostics fabricated with widely available resources.
In the following section, we briefly discuss the mechanisms of amplification and visual discrimination between positive and negative samples.
Colorimetric LAMP amplification
The presence of phenol red within the LAMP reaction mix allows for naked-eye discrimination between positive and negative samples. The reaction mix is coupled with the pH color transition of phenol red, a widely used pH indicator, which shifts in color from red to yellow at pH 6.8. During LAMP amplification, the pH of the reaction mix continuously evolves from neutrality to acidic values as protons are produced [27,47]. The mechanism of production of hydrogen ions (H+) during amplification in weakly buffered solutions has been described [47]. DNA polymerases incorporate a deoxynucleoside triphosphate into the nascent DNA chain. During this chemical event, a pyrophosphate moiety and a hydrogen ion are released as byproducts (Figure 2 A). This release of hydrogen ions is quantitative, according to the reaction scheme illustrated in Figure 2. The caudal of H+ is high, since it is quantitatively proportional to the number of newly integrated dNTPs. In fact, the quantitative production of H+ is the basis of previously reported detection methods, such as the semiconductor sequencing technology operating in Ion Torrent sequencers[48].
In the initially neutral and weakly buffered reaction mixes, the production of H+ during LAMP amplification progressively and rapidly shifts the pH across the threshold of phenol red (Figure 2B).
Moreover, the pH shift is clearly evident to the naked eye, thereby freeing the user from reliance on spectrophotometric instruments and facilitating simple implementation during emergencies (Figure 2C). Images in Figure 2C show representative colors of the amplification reaction mixes contained in Eppendorf PCR tubes after incubation for 30 min. Three different incubation temperatures were tested (50, 60, and 65 °C) and two different sets of LAMP-primers (α and β) were used.
Both sets of primers performed equivalently, at least based on visual inspection, in the three temperature conditions tested. Discrimination between positive and negative controls is possible using only the naked eye to discern the reaction products from amplifications conducted at 60 and 65 °C. No or negligible amplification was evident at 50 °C or in the control group.
Furthermore, we were able to successfully discriminate between positive and negative samples using LAMP reaction mix already added with primers and kept at room temperature for 48 h or at 4 °C for 72 hours (Figure S3). The stability of the reaction, the isothermal nature of the amplification process, and its independence from specialized equipment greatly simplifies the logistic of implementation of this diagnostic method outside centralized labs.
Analysis of sensitivity
We conducted a series of experiments to assess the sensitivity of the LAMP reactions in the 3D-printed incubation chamber using the two sets of primers (α and β; Table 1). The amplification proceeds with sufficient quality to also allow proper visualization of the amplification products in electrophoresis gels, even at low nucleic acid concentrations. We observed that amplification proceeded successfully in a wide range of viral loads, from 625 to 5 × 105 copies in experiments using synthetic SARS-CoV-2 nucleic acid material (Figure 3A).
We clearly observed amplification in samples containing as few as 625 viral copies after incubation times of 5 min at 65 °C. If we put this range into a proper clinical context, the actual viral load of COVID-19 in nasal swabs from patients has been estimated to fall within the range of 105 to 106 viral copies per mL [49]. Discrimination between positive and negative samples (controls) can be clearly established by the naked eye in all reactions incubated for 50 min, regardless of the number of viral copies present. In addition, we did not observe any non-specific amplification in negative samples (i.e., containing synthetic genetic material form EBOV) incubated for 50 min at 65 °C. Indeed, the identification and amplification of SARS-CoV-2 synthetic material is feasible in samples that contained ∼62.5 viral copies using this LAMP strategy (Figure S3) and incubation times of 50–60 min.
We corroborated the amplification by visualizing LAMP products with gel electrophoresis for the different viral loads tested. Figures 3B,C show agarose gels of the amplification products of each one of the LAMP experiments, where two different sets of primers (α and β) were used to amplify the same range of concentrations of template (from 625 to 2 × 105 synthetic viral copies). We were able to generate a visible array of bands of amplification products, a typical signature of LAMP, for both LAMP primer sets and across the whole range of synthetic viral loads. Indeed, both primer sets rendered similar amplification profiles.
In summary, using the primers and methods described here, we were able to consistently detect the presence of SARS-CoV-2 synthetic nucleic acids. We have used a simple 3D-printed incubator, connected to a water circulator, to conduct LAMP. We show that, after only 30 min of incubation, samples containing a viral load in the range of 104 to 105 copies could be clearly discriminated from negative samples by visual inspection with the naked eye (Figure 2C). Samples with a lower viral load were clearly discriminated when the LAMP reaction was incubated for 50 min. Incubation periods of up to 1 h at 68 °C did not induced false positives and were able to amplify as few as ∼62 copies of SARS-CoV-2 synthetic genetic material. These results are consistent with those of other reports in which colorimetric LAMP, assisted by phenol red, has been used to amplify SARS-COV-2 genetic material [39,40].
We observe 0 false positive cases in experiments where synthetic samples containing EBOV genetic material were incubated at 65 °C for 1 h.
In the current context of the COVID-19 pandemics, the importance of communicating this result does not reside in its novelty but in its practicality. Some cost considerations follow. While the market value of a traditional RT-qPCR apparatus (the current gold standard for COVID-19 diagnostics) is in the range of 10,000 to 40,000 USD, a 3D-printed incubator, such as the one described here (Figure S1,S2; Supplementary file S1), could be fabricated for under 200 USD at any 3D printer shop. This difference is significant, especially during an epidemic or pandemic crisis when rational investment of resources is critical. While the quantitative capabilities of testing using an RT-qPCR platform are undisputable, the capacity of many countries to rapidly, effectively, and massively establish diagnostic centers based on RT-qPCR is questionable. The current pandemic scenarios experienced in the USA, Italy, France, and Spain, among others, have crudely demonstrated that centralized labs are not an ideal solution during emergencies. Portable diagnostic systems may provide a vital flexibility and speed of response that RT-qPCR platforms cannot deliver.
Feasibility of real-time quantification
Here, we further illustrate the deterministic and quantitative dependence between the concentration of the amplification product and the color signal produced during this colorimetric LAMP reaction. For this purpose, we simulated real-time amplification experiments by conducting a series of amplification reactions using initial amounts of 625, 1 × 104, and 2 × 105 copies of synthetic SARS-CoV-2 genetic material in our 3D-printed incubator.
We extracted samples from the incubator after 0, 10, 20, 30, 40, and 50 min of incubation at 65 °C. The color of these samples was documented as images captured using a smart phone (iPhone 7) against a white background (Figure 4A). The images were analyzed using the free access application Color Companion® for the iPhone or iPad. Briefly, color images were decomposed into their CIELab space components. In the CIELab color space, each color can be represented as a point in a 3D-space, defined by the values L, a, and b [45]. In this coordinate system, L is the luminosity (which ranges from 0 to +100), a is the blue-yellow axis (which ranges from -50 to 50), and b is the green-red axis (which ranges from - 50 to 50) (Figure S4).
The difference between two colors can be quantitatively represented as the distance between the two points that those colors represent in the CIELab coordinate system. For the colorimetric LAMP reaction mixture used in our experiments, the spectrum of possible colors evolves from red (for negative controls and negative samples) to yellow (for positive samples). Conveniently, the full range of colors for samples and controls can be represented in the red and yellow quadrant defined by L [0,50], a [0,50], and b [0,50]. For instance, the difference between the color of a sample (at any time of the reaction) and the color of the negative control (red; L=53.72 ± 0.581, a=38.86 ± 2.916, and b=11.86 ± 0.961) can be calculated in the CIELab space.
We determined the distance in the CIELab space between the color of samples taken at different incubation times that contained SARS-CoV-2 genetic material and negative controls (Figure 4B and C). We repeated this calculation for each of the LAMP primer sets that we used, namely primer set α (Figure 4B) and β (Figure 4C).
These results suggest that the color difference between the samples and negative controls is quantifiable. Therefore, color analysis may be implemented to assist the discrimination between positives and negatives. Furthermore, imaging and color analysis techniques may be implemented in this simple colorimetric LAMP diagnostic strategy to render a real-time quantitative Lamp (RT-qLAMP). Alternatively, the progression of the amplification at different times can be monitored by adding an intercalating DNA agent, (i.e., EvaGreen Dye), and measuring fluorescence on time (Figure S5).
Note that the variance coefficients for the control are 1.08, 7.50, and 8.10% for L, a, and b, respectively. These small values suggest robustness and reproducibility in the location of the coordinates of the control point (reference point). Similarly, the variation in color between negative controls and positive samples incubated for 50 min was reproducible and robust (average of 46.60 +/-4.02 d.u.; variance coefficient of 8.62%).
Finally, we observed differences in the performance of the two LAMP primer sets used in the experiments reported here (Figure 4B and 5).
Our results suggest that primer set α enabled faster amplification in samples with fewer viral copies. Consistently, this primer set yielded positive discrimination in samples with 625 viral copies in 30 min (Figure 4B). The use of primer set β enabled similar differences in color, measured as distances in the CIELAB 3D-space, in 40 min (Figure 4C). These findings suggest that primer set α should be preferred for final-point implementations of this colorimetric LAMP method. Interestingly, primer set β may better serve the purpose of a real-time implementation. While primer set α produced similar trajectories of evolution of color in samples that contained 1.0 × 104 and 2.0 × 105 viral copies (Figure 5A), primer set β was better at discriminating between amplifications produced from different initial viral loads (Figure 5B).
Conclusions
The challenge of point-of-care detection of viral threats is of paramount importance, particularly in underdeveloped regions and in emergency situations (i.e., epidemic outbreaks). In the context of the current COVID-19 pandemic, the availability of testing infrastructure based on RT-qPCR is recognized as a serious challenge around the world. In developing economies (i.e. Latin America, India, and most African countries), the currently available resources for massive COVID-19 testing by RT-qPCR will clearly be insufficient. Even in developed countries, the time to get diagnostic RT-qPCR results from a COVID-19 RT-qPCR test currently ranges from 1 to 5 days. Clearly, the available PCR labs are overburdened with samples, have too few personnel to conduct the tests, are struggling with backlogs on the instrumentation, and face complicated logistics to transport delicate and infective samples while preserving the cold chain.
Here, we have demonstrated that a simple embodiment of a LAMP reaction, assisted by the use of phenol red as a pH indicator and the use of a simple 3D-printed chamber connected to a water circulator, can enable the rapid and highly accurate identification of samples that contain artificial SARS-CoV-2 genetic sequences. Amplification is visually evident, without the need for any additional instrumentation, even at low viral copy numbers. In our experiments with synthetic samples, we observed 100 percent accuracy in samples containing as few as 625 copies of SARS-CoV-2 genetic material.
Validation of these results using real human samples from positive and negative COVID-19 subjects is obviously needed to obtain a full assessment of the potential of this strategy as an alternative to RT-qPCR platforms. However, our results with synthetic samples suggest that this simple strategy may greatly enhance the capabilities for COVID-19 testing in situations where RT-qPCR is not feasible or is unavailable.
Data Availability
All raw data related to the research described in this manuscript has been made available as supplemental material.
Supporting Information
Figure S1. (A) Actual images, and (B) rendering of the 3D-printed incubation chamber used in the LAMP experiments.
Figure S2. Schematic representation of the chamber (different views) showing its relevant dimensions.
Figure S3. (A) The colorimetric LAMP method described here was able to identify and amplify synthetic SARS-CoV-2 genetic material in samples containing as few as ∼62 viral copies. (B) Evaluation of the stability and functionality of the LAMP reaction mix at different storage times and temperatures. The reaction mix, which is formulated with LAMP primers and ready for the addition of nucleic acid extracts, is functional and discriminates between positive and negative samples when stored (i) at room temperature for 48 h or (ii) at 4 °C for 72 h.
Figure S4. (A) Color analysis conducted on positive and negative SARS-CoV-2 samples contained in Eppendorf PCR tubes (yellow inset) using Color Companion, a freely available app from Apple (downloadable at Apple Store, USA). This app identifies the components of color in a specific location of an image (black circle in the yellow inset) in the CIELab, RGB, HSB, or CMYK spaces. The image can be uploaded using e-mail, airdrop, or Whatsapp. (B) Schematic representation of the CIELab space, a color system where any color can be represented in terms of a point and its coordinates in a 3D space, where L is luminosity, a is the axis between green and red, and b is the axis between yellow and red.
Figure S5. (A) The amount of amplification product in LAMP experiments was evaluated by measuring the fluorescence emitted by the amplification product in reactions with an added intercalating agent. Fluorescence readings were conducted in standard 96-well plates and using a conventional plate reader. (A) Fluorescence readings, as measured in a commercial plate reader, for different dilutions of SARS-CoV-2 synthetic DNA templates. Results using two different LAMP primer sets are shown: set α (indicated in blue), and set β (indicated in red).
Acknowledgments
EGG acknowledges funding from a doctoral scholarship provided by CONACyT (Consejo Nacional de Ciencia y Tecnología, México). GTdS and MMA acknowledge the institutional funding received from Tecnológico de Monterrey (Grant 002EICIS01), and funding provided by CONACyT (Consejo Nacional de Ciencia y Tecnología, México) through grants (SNI 26048, SNI 256730, and Scholarships 635891, 856068, and 814593).