Abstract
Objective To implement parallel transmit (pTx) 7T MRI in the pre-surgical evaluation of patients with drug resistant focal epilepsy, and to compare quality and diagnostic yield to conventional single transmit (specifically, circularly polarised, CP) 7T MRI.
Methods We implemented a comparative protocol comprising both pTx and CP 7T MRI in consecutive adult candidates for epilepsy surgery who had negative or equivocal 3T MRI imaging. Here we report the outcomes from the first 31 patients.
We acquired pTx and CP T1, T2, FLAIR and EDGE images, all in the same 3D 0.8mm isotropic space. 2D high-resolution T2 and T2*-weighted sequences were acquired only in CP mode due to current technological limitations.
Two neuroradiologists, a neurologist and a neurosurgeon made independent, blinded quality and preference ratings of pTx vs CP images. Quantitative methods were used to assess signal dropout.
Results Blinded comparison confirmed significantly better overall quality of pTx FLAIR images (F(2,184)=13.7, p=2.88×10-6), while pTx MP2RAGE images were subjectively non-inferior and had improved temporal lobe coverage with quantitatively less signal drop-out.
7T-pTx revealed previously-unseen structural lesions in 9 patients (29%), confirmed 3T-equivocal lesions in 4 patients (13%), and disproved 3T-equivocal lesions in 4 patients (13%).
Lesions were better visualised on pTx than CP in 57% of cases, and never better visualised on CP.
Clinical management was altered by pTx-7T in 18 cases (58%). 9 cases were offered surgical resection and 1 LITT. 3 cases were removed from the surgical pathway because of bilateral or extensive lesions. 5 cases were offered sEEG with better targeting (in 3 because the 7T lesion was deemed equivocal by the MDT, and in 2 because the lesion was extensive).
Significance Parallel transmit 7T MRI is implementable in a clinical pathway, is superior to single transmit 7T MRI, and changed management in 58% of patients scanned.
Key points
We scanned 31 patients with parallel transmit and conventional 7T MRI, finding previously-unreported structural lesions in 9 patients (29% of cases).
In 13% of cases pTx 7T MRI showed that an equivocal lesion at 3T MRI was likely significant.
In 13% of cases pTx 7T MRI showed that an equivocal lesion at 3T MRI could be disregarded.
Both qualitative and quantitative quality assessments indicate superiority of pTx images over CP.
Future clinical implementations of 7T MRI for epilepsy should utilise parallel transmit where possible.
Competing Interest Statement
CTR discloses research grant support from Siemens, for a different project. AM is employed by Siemens Healthcare SAS, Saint-Denis, France The remaining authors have no conflicts of interest.
Clinical Trial
ISRCTN16452095
Funding Statement
This work was supported by the Cambridge University Hospitals Academic Fund for functional neurosurgery and the University of Cambridge Medical Research Council Impact Acceleration Account Award (UKRI Grant Reference - MR/X502844/1). MZ was supported by an MRC PhD studentship (MR N013433-1). This study was supported by the NIHR Cambridge Biomedical Research Centre (NIHR203312) and an MRC Clinical Research Infrastructure Award for 7T (MR/M008983/1).
Author Declarations
I confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.
Yes
The details of the IRB/oversight body that provided approval or exemption for the research described are given below:
The UK National Research Ethics Service West Midlands - Solihull Research Ethics Committee gave ethical approval for this work (23/WM/0008)
I confirm that all necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived, and that any patient/participant/sample identifiers included were not known to anyone (e.g., hospital staff, patients or participants themselves) outside the research group so cannot be used to identify individuals.
Yes
I understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).
Yes
I have followed all appropriate research reporting guidelines, such as any relevant EQUATOR Network research reporting checklist(s) and other pertinent material, if applicable.
Yes
Footnotes
↵* Joint senior authors.
Data Availability
Scan data are available on reasonable request but not publicly available due to patient confidentiality.
Abbreviations
- CP
- Circularly Polarized acquisition (equivalent to sTx using the parallel transmit coil)
- DTI
- Diffusion Tensor Imaging
- EDGE
- Edge-Enhancing Gradient Echo (first inversion image from MP2RAGE)
- FCD
- Focal Cortical Dysplasia
- FLAIR
- Fluid Attenuated Inversion Recovery
- ILAE
- International League Against Epilepsy
- MDT
- Multi Disciplinary Team
- NAAD
- Normalised Absolute Average Deviation
- pTx
- parallel transmit acquisition
- sEEG
- stereotactic Electroencephalography
- sTx
- single transmit acquisition
- UHF
- Ultra High Field
- UNI
- MP2RAGE uniform image