ABSTRACT
Background Chirp is a privacy-preserving radar sensor developed to continuously monitor older adults’ safety and mobility without the need for cameras or wearable devices. Our study purpose was to evaluate the inter-sensor reliability, intrasession test-retest reliability, and concurrent validity of Chirp in a clinical setting.
Methods We recruited 35 community-dwelling older adults (mean age 75.5 (standard deviation: 6.6) years, 86% female). All participants lived alone in an urban city in southwestern Ontario and had access to a smart device with wireless internet. Data were collected with a 4-meter ProtoKinetics Zeno™ Walkway (pressure sensors) with the Chirp sensor (radar positioning) at the end of the walkway. Participants walked during normal and adaptive locomotion experimental conditions (walking-while-talking, obstacle, narrow walking, fast walking). Each of the experimental conditions was conducted twice in a randomized order, with fast walking trials performed last. For intra-session reliability testing, we conducted two blocks of walks within a participant session separated by approximately 30 minutes. Intraclass Correlation Coefficient(A,1) (ICC(A,1)) was used to assess the reliability and validity. Linear regression, adjusted for gender, was used to investigate the association between Chirp and cognition and health-related quality of life scores.
Results The Chirp inter-sensor reliability ICC(A,1)=0.999[95% Confidence Interval [CI]: 0.997 to 0.999] and intrasession test-retest reliability [ICC(A,1) =0.921, 95% CI: 0.725 to 0.969] were excellent across all experimental conditions. Chirp concurrent validity compared to the ProtoKinetics Zeno™ Walkway was excellent across experimental conditions [ICC(A,1)= 0.993, 95% CI: 0.985 to 0.997]. We found a weak association between Chirp and cognition scores using the Montreal Cognitive Assessment across experimental conditions (estimated β-value= 7.79, 95% CI: 2.79 to 12.80) and no association between the Chirp and health-related quality of life using the 12-item Short Form Survey across experimental conditions (estimated β-value=6.12, 95% CI: -7.12 to 19.36).
Conclusion Our results demonstrate that Chirp is a reliable and valid measure to assess gait parameters in clinics among older adults.
Competing Interest Statement
The authors have declared no competing interest.
Funding Statement
Yes
Author Declarations
I confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.
Yes
The details of the IRB/oversight body that provided approval or exemption for the research described are given below:
Ethics approval was obtained from the Hamilton Integrated Research Ethics Board (HIREB #15237).
I confirm that all necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived, and that any patient/participant/sample identifiers included were not known to anyone (e.g., hospital staff, patients or participants themselves) outside the research group so cannot be used to identify individuals.
Yes
I understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).
Yes
I have followed all appropriate research reporting guidelines, such as any relevant EQUATOR Network research reporting checklist(s) and other pertinent material, if applicable.
Yes
Data Availability
See file S4 in the supplementary information