Abstract
Objective Among the advancements in computed tomography (CT) technology, photon-counting computed tomography (PCCT) stands out as a significant innovation, providing superior spectral imaging capabilities while simultaneously reducing radiation exposure. Its long-term stability is important for clinical care, especially longitudinal studies, but is currently unknown. This study sets out to comprehensively analyze the long-term stability of a first-generation clinical PCCT scanner.
Methods Over a two-year period, from November 2021 to November 2023, we conducted weekly identical experiments utilizing the same multi-energy CT protocol. Throughout this period, notable software and hardware modifications were meticulously recorded. Various tissue-mimicking inserts were scanned weekly to rigorously assess the stability of Hounsfield Units (HU) and image noise in Virtual Monochromatic Images (VMIs) and iodine density maps.
Results Spectral results consistently demonstrated the quantitative stability of PCCT. VMIs exhibited stable HU values, such as variation in relative error for VMI 70 keV measuring 0.11% and 0.30% for single-source and dual-source modes, respectively. Similarly, noise levels remained stable with slight fluctuations linked to software changes for VMI 40 and 70 keV that corresponded to changes of 8 and 1 HU, respectively. Furthermore, iodine density quantification maintained stability and showed significant improvement with software and hardware changes, especially in dual-source mode with nominal errors decreasing from 1.44 to 0.03 mg/mL. Conclusion This study provides the first long-term reproducibility assessment of quantitative PCCT imaging, highlighting its potential for the clinical arena.
Key Points Question: Photon-counting CT (PCCT) provides critical spectral imaging for improved diagnostic accuracy, but its long-term quantitative stability over time is still unknown.
Findings: The clinical PCCT system demonstrated stable Hounsfield Units (HU) and image noise over two years, ensuring reliable quantitative imaging and improving diagnostic accuracy.
Clinical Relevance: This study showcased the exceptional value of PCCT in diagnostic radiology, particularly for its application in longitudinal studies.
Competing Interest Statement
PN and HI have a research agreement with Siemens Healthineers. PS is an employee of Siemens Medical Solutions.
Funding Statement
We acknowledge support from National Institutes of Health (R01EB030494) and Siemens Healthineers.
Author Declarations
I confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.
Yes
I confirm that all necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived, and that any patient/participant/sample identifiers included were not known to anyone (e.g., hospital staff, patients or participants themselves) outside the research group so cannot be used to identify individuals.
Yes
I understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).
Yes
I have followed all appropriate research reporting guidelines, such as any relevant EQUATOR Network research reporting checklist(s) and other pertinent material, if applicable.
Yes
Footnotes
The manuscript has been updated to its final peer-reviewed version.
Data Availability
All data produced in the present study are available upon reasonable request to the authors.
Abbreviations
- CT
- computed tomography
- TB
- tuberculosis
- PCCT
- photon-counting CT
- VMI
- virtual monoenergetic images
- HU
- Hounsfield units
- QIR
- quantum iterative reconstruction
- ROI
- region of interest
- PELT
- pruned exact linear time