ABSTRACT
Background Polysomnography (PSG) is the gold standard for sleep monitoring and diagnosis, yet it is difficult to use in home environments. This study evaluated the performance of a wearable electroencephalographic (EEG) device, LANMAO Sleep Recorder in EEG recording and sleep staging algorithm by comparing with PSG.
Method Sleep of 7 Chinese adults were recorded concurrently with PSG and LANMAO devices. First, we validated the consistency of the raw signals with relative spectral power and Pearson correlation coefficient. Second, we evaluated the performance of the automated sleep staging algorithm integrated in the LANMAO device by comparing with the staging by experts.
Results The Pearson correlation coefficient between the relative spectral power of multiple frequency bands during the sleep stages ranged from 0.7613 to 0.8816, with the strongest correlation observed for delta waves (r=0.8816). The overall F1-Score of the automated sleep staging algorithm was 84.03%, with individual F1-Scores for each class as follows: Wake: 93.67%, REM: 87.23%, Light Sleep: 72.10%, and Deep Sleep: 82.82%.
Conclusion The results suggest that the EEG recorded by the LANMAO Sleep Recorder is precise and valid, and its automated sleep staging algorithm can accurately perform sleep staging with high accuracy. Therefore, in specific scenarios such as the home environment, LANMAO devices can work as a promising PSG alternative for sleep monitoring.
Competing Interest Statement
This study was supported by Shanghai Quanlan Technology team who provided the wearable EEG recording devices.
Funding Statement
This study did not receive any funding.
Author Declarations
I confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.
Yes
The details of the IRB/oversight body that provided approval or exemption for the research described are given below:
All procedures in this study were approved by the IRB, Shanghai Mental Health Center (2022-12RI).
I confirm that all necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived, and that any patient/participant/sample identifiers included were not known to anyone (e.g., hospital staff, patients or participants themselves) outside the research group so cannot be used to identify individuals.
Yes
I understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).
Yes
I have followed all appropriate research reporting guidelines, such as any relevant EQUATOR Network research reporting checklist(s) and other pertinent material, if applicable.
Yes
Data Availability
All data produced in the present study are available upon reasonable request to the authors.