Abstract
Objective In this research a non-invasive ICP optical sensor has been developed and evaluated in a clinical pilot study. The technology is based on infrared light interrogating brain tissue, including photodetectors used to detect the backscattered light, which is modulated by vascular pulsations in the brain vascular tissue. The hypothesis underpinning this research is based on the expected changes in the extramural arterial pressure affecting the morphology of the recorded optical signals (photoplethysmograms (PPGs)), therefore analysis of the acquired signals using a bespoke algorithm could enable the calculation of the intracranial pressure non-invasively (nICP).
Methods This pilot study is the first evaluation of the nICP probe in patients in whom the gold standard comparator of invasive ICP was available. nICP monitoring was performed for up to 48 hours in each of the 40 patients undergoing invasive ICP monitoring as part of their prescribed medical treatment. The quality of the recorded PPG signals was analysed, and time-based features were extracted off-line. Data from all ICP levels were randomly allocated into two groups to train (80%) and test (20%) a machine learning algorithm. A Bland Altman analysis and ROC curve were calculated to evaluate the accuracy of the estimated nICP value compared to the gold standard of synchronously acquired invasive ICP data.
Results The successful acquisition of cerebral PPG signals from TBI patients for the subsequent extraction of morphological features allowed the generation of a bagging tree model to estimate ICP non-invasively. The non-invasive estimation of ICP was achieved with 95% limits of agreement of 3.8 mmHg and a negligible bias. Furthermore, the model achieved a good correlation coefficient of 82.54% compared to standard clinical invasive ICP monitoring. Finally, the ROC curve analysis showed good diagnostic capability with a sensitivity of 80% and specificity of 89%.
Conclusion Clinical evaluation of this novel optical nICP sensor demonstrated for the first time that it could estimate ICP non-invasively to an acceptable and clinically useful accuracy and paves the way for further technology optimisation and larger clinical studies.
Introduction
Traumatic Brain Injury (TBI) is a global pandemic affecting 50 million patients globally each year with an annual cost to the global economy of £400 billion1. It is the most common cause of death and disability in the under-40 age group, and its incidence is increasing2. Secondary brain injury due to raised intracranial pressure (ICP) may result in progressive cerebral ischaemia, herniation syndromes and death at various raised ICP thresholds3,4. ICP monitoring is used in unconscious patients to guide treatment5, and despite a trial where ICP monitoring was compared to imaging and clinical examination (ICE), which showed no difference in outcomes6, ICP-directed therapy remains the international recommended standard of care for severe TBI7. A European survey subsequently found that 90% of clinicians would insert an ICP monitor in patients with severe TBI and radiological abnormalities8, and a recent meta-analysis suggested that ICP directed therapy was associated with a lower mortality9.
All currently available ICP monitoring systems are invasive and require access to the cranial cavity, which have a small but significant risk of complications, including haemorrhage or infection10, and for this reason they are generally only inserted by neurosurgeons. The quest for a non-invasive monitor has the immediate advantage that it will eliminate the risk of complications associated with the invasiveness and obviate the requirement for a neurosurgeon to insert it.
Other techniques have been investigated to measure ICP non-invasively11–15. These include, Transcranial Doppler (TCD) based techniques16–21 or the measurement of the Optic Nerve Sheet Diameter (ONSD)22,23 .However, these devices involve cumbersome technology, are operator-dependent, and may not provide direct and continuous measurement of ICP24. Also, none of these technologies have been accepted in routine clinical practice for measuring ICP, and hence their clinical efficacy and accuracy has not been extensively demonstrated yet24.
In the UK, patients with severe TBI requiring time-critical surgery can take up to 6 hours to reach a neurosurgeon if initially resuscitated in a hospital without a neurosurgery department25 and 3.7 hours if taken straight to a neurosurgery hospital26. In contrast, a non-invasive ICP monitor used by pre-hospital emergency services would allow ICP-directed therapy (e.g., osmodiuretics, deep sedation and hyperventilation) to be implemented as soon as the patient is reached, certainly within the “golden hour” and at a far earlier stage than what is currently achievable in most situations. Therefore, there is a significant need to develop novel technologies that will allow truly non-invasive and continuous ICP measurements which are neither cumbersome nor operator-dependent.
The technique of near-infrared spectroscopy (NIRS) has been extensively used to assess cerebral oxygenation non-invasively; however, current NIRS devices only provide relative changes of oxy-, deoxy-haemoglobin and tissue oxygenation index (Hb, HbO2, TOI) and only use non-pulsatile brain signals. To-date, there is no evidence of any non-invasive technology that uses the pulsatile component of near-infrared signals to assess ICP quantitatively14. Therefore, this research utilised a non-invasive, continuous monitoring system to acquire cerebral pulsatile NIRS signals (Photoplethysmograms (PPGs)), referred as PPG-NIRS, from the forehead of TBI patients. Such technology was developed with the aim of allowing ICP monitoring early and in a variety of settings, thereby decreasing the risks of secondary injury and reducing costs. The developed sensor works by shining infrared light into the brain through the skull, and records PPG signals from the backscattered light detected by two photodetectors.
This pilot study is based on the hypothesis that changes in the extramural arterial pressure (intracranial hypertension) will affect the morphology of the cerebral PPG signals, so advanced algorithms and Machine Learning (ML) models utilising optical signal feature extraction techniques could be implemented in translating the PPG-NIRS signals into absolute measurements of ICP within acceptable accuracy limits. The results of this pilot clinical investigation will guide future device optimisation and the design of subsequent clinical studies.
Middle section
Methods
Study design
This 78-week, non-randomised pilot study (ClinicalTrials No. NCT05632302) was performed at a single site in the United Kingdom from January 2020 to July 2021 (the study was delayed significantly by the COVID-19 global pandemic). The East of England - Cambridge Central Research Ethics Committee approved the protocol on 14/02/19 (REC reference 18/EE/0276, IRAS ID 219476). The Medicines and Healthcare Products Regulatory Agency (MHRA) had no grounds for objection to making the device available for the purposes of a clinical investigation (CI/2019/0025). This single-arm trial recorded optical signals using the interventional device, while invasive intracranial pressure measurements were recorded simultaneously using a traditional invasive method as part of their normal medical treatment. This study was performed in accordance with the Declaration of Helsinki and in agreement with the International Conference on Harmonisation Guidelines on Good Clinical Practice. This manuscript was written following the preferred reporting items presented on the “CONSORT 2010 checklist of information to include when reporting a pilot or feasibility trial” 27.
Participants
Participants were recruited on the intensive therapy unit (ITU) at the Royal London Hospital, UK on the advice of a personal or professional consultee. Severe TBI diagnoses were made based on guidelines for the Management of Severe Traumatic Brain Injury, Fourth Edition 7. With the exemption of one patient, all the subjects were unconscious as a result of head injury, which was the indication of invasive ICP monitoring. One conscious patient undergoing invasive ICP monitoring as an investigation for normal pressure hydrocephalus was also included in the study.
Potential participants were excluded if they were deemed unlikely to survive 48 hours or if a personal consultee advised against participation. Patients who had undergone decompressive craniectomy were excluded due to poor signal quality due to the damping effect of the surgery.
All conscious participants provided written informed consent prior to study participation. In the case patients who could not consent for themselves due to incapacity due to TBI, consent was obtained from a personal consultee or an independent healthcare professional if a personal consultee was unavailable, according to the 2005 UK Mental Capacity Act.
Personal information relating to participants has been kept confidential and managed under the Data Protection Act, NHS Caldecott Principles, The Research Governance Framework for Health and Social Care, and the conditions of Research Ethics Committee Approval.
Interventions
The reference invasive ICP monitor used was the Raumedic® Neurovent-P intra-parenchymal pressure probe43 (catalogue reference 092946001) and was used in all patients. This was inserted through a single-lumen 5-French polymer cranial bolt and interfaced with a GE® monitor. The reference invasive ICP was collected via the GE® iCollect software on a laptop computer powered by a medical-grade isolation transformer (SN: 2430).
The developed reflectance non-invasive optical ICP sensor(nICP) was placed on the patient’s forehead, and optical signals were acquired from extracerebral tissues and the brain (Figure 1). The nICP sensor comprised a high-intensity light-emitting diode (OIS-330-810-X, Osaoptolight, Germany) with a peak emission wavelength of 810 nm and two silicone photodiode detectors with a large active area (VBPW34S, Vishay Intertechnology, USA) positioned at 10 mm and 35 mm from the light source28. These distances were chosen to guarantee the light penetration necessary to interrogate extracerebral tissues and the cerebral cortex respectively, as is commonly used in cerebral Near Infrared Spectroscopy measurements29,30.
The probe was connected to a custom-made instrumentation unit responsible for supplying the driving currents to the LED, transforming and amplifying light intensities into voltages and pre-processing the acquired signals. This system was previously described as the ZenPPG31, and it was powered by a battery pack. From the ZenPPG, the signals passed through a National Instruments Data Acquisition Card (DAQ) card to a laptop, where an in-house Labview acquisition software recorded the data sampled at 100Hz.
After the invasive ICP monitor had been inserted, monitoring commenced. The nICP monitor was calibrated in each case, adjusting the LED intensity and amplification gain according to patients’ characteristics and ambient light. Calibration was performed before recording started, then monitoring continued for up to 48 hours, and the investigators monitored the signals during this time. If the patient left the ITU for a scan or surgery, the nICP monitor probe was disconnected but left in situ.
Outcomes
Endpoints, assessment measurements, and the criteria to proceed with a further clinical trial are summarised in Table 1.
The primary endpoint was the accuracy of the nICP model calculated from the acquired optical signals. Moreover, key secondary endpoints were the quality of the optical signals and safety issues related to the probe. Outcomes were evaluated at the end of data collection and complete model analysis for the primary and key secondary endpoints. The nICP model accuracy and the probe’s safety included prespecified criteria used to judge whether or how to proceed with future clinical trials using the nICP monitor.
Sample size
The GE® monitor uses a sampling frequency of 100 Hz to acquire invasive ICP measurements. Intending to keep the nICP signals consistent with the invasive data, the sampling frequency defined in the LabView interface for the optical signals acquisition was also at 100 Hz. In addition, a monitoring window of up to 48 hours was predefined based on published literature that suggests a range of elective ICP monitoring between 8 hours and 72 hours32. Moreover, the selected monitoring window considered that ICP monitoring devices are generally removed once the ICP is normalised with sustained or clinical improvement for at least 48 hours33. The features from the optical signals will be explained in the following statistical methods section. Features were extracted from the PPG signals every 60 seconds, leading to a record of up to 2880 observations per patient for each feature.
It remains unclear how to calculate the sample size required for a given machine learning model with the aim of medical application34. Therefore, the patients’ sample size was calculated based on a predefined desired nICP model’s diagnostic accuracy. Since the accuracy reported for other non-invasive ICP monitoring techniques35 reaches 90% for both sensitivity and specificity, 90% was established as the nICP model’s desired diagnostic accuracy. According to Headway (UK national brain injury charity) statistical reports, the incidence of TBI admission in London hospitals was 216 per 100,000 of the population between 2019 and 2020. Given these assumptions and defining α=0.1 and β=0.2, the sample size was calculated as 37 patients. 10% was added due to potential attrition therefore the final recruitment was 40 patients.
In accordance with the referenced checklist published by ClinicalTrials.gov36, this study cannot be considered a small clinical trial despite the intentions of the authors to test a prototype device. Where the primary outcome measure relates to feasibility and not to health outcomes (the latter defined as a trial with at least 10 subjects), the current study would generally not be considered “small” for purposes of the exclusion36.
Randomisation, Implementation and Blinding
Considering that this is a single-arm study, where both nICP and invasive monitoring occurred synchronously in all the patients, there was no intervention randomisation. However, during the off-line analysis of the optical signals, data were randomly assigned to the training or the testing group.
The function cvpartition of MATLAB was utilised to randomly allocate the dataset to the training group (80%) and the testing group (20%), following the Pareto law37. Because most patients (39/40) were receiving pharmacological sedation with the aim of maintaining an ICP of less than 20mmHg, the distribution of the ICP across the whole dataset is biased toward normal ICP. Therefore, the dataset was divided by ICP levels (5-10mmHg, 10-15 mmHg, …,35-40 mmHg) before applying the partition function to each interval. This ensured that data from all ICP levels were included in both training and testing groups. Due to the nature of the study no blinding was employed. Patients were numbered sequentially by recruitment date using natural numbers starting from number one (1). A key linking these sequential numbers to patient details was stored on a secure clinical computer however other than this, there is no way to track the patient’s personal information.
Statistical methods
General signal processing was performed using Matlab R2022b (The MathWorks Inc., Natick, MA, USA). All PPG signals are sensitive to different noise sources, such as movement and ambient light. Therefore, movement artefact anomalies and photodetector saturation sections were removed from the cerebral PPG signals, as well the respective synchronous segment was removed from the invasive ICP reference. Then, the cerebral PPG signals were filtered using Butterworth filters to separate the AC PPG component (2nd order bandpass filter with cut-off frequencies of 0.8 and 10 Hz) from the DC PPG component (2nd order lowpass filter with a cut-off frequency of 0.1 Hz). The filtered signals were then normalised by dividing the AC part of the signal by its DC, followed by a 10-factor multiplication.
Several features have been extracted and investigated in the literature to characterise pulsating signals, such as PPG38. This study extracted eleven time-domain features from the cerebral PPG signals. These features were:
amplitude
pulse width
rise time
decay time
upslope
downslope
area under the curve
area of the systolic period
area of the diastolic period
ratio between both systolic and diastolic areas
ratio between the max and min of the second derivative pulse.
The median value of each feature was calculated during a signal window of 60 seconds. Also, the mean value of invasive ICP was calculated for this period. Given the sparsity of recorded pulses above 20 mmHg of ICP, the final dataset only included ICP values between 5 and 40 mmHg. The above procedures resulted in a total of 40,795 observations across all patients.
As mentioned, the dataset of observations (features + mean ICP / window) was randomly portioned in training and testing sets. An additional set (20%) was held out from the training group for cross-validation of the model parameters; this group is called the validation set. Both training and validation sets were used to build a regression bagged tree model using the regression learner app of Matlab R2022b39. The functionality and working principle of regression trees and bagging are explained in detail elsewhere 40. The regression chosen here was a bagged ensemble of 30 individual decision tree regressors. The maximum depth of each tree was set to 8, considering that an increased tree depth can yield better results but risks overfitting. Also, no restrictions were given on the features available to every node or tree, so all features could be used at any time.
After training the regression forest with the training data set and the corresponding invasively measured ICP as the ground truth, the testing data set was used. Bland-Altman analysis was performed to determine the inter-method agreement. The correlation between methods was assessed by considering the Pearson correlation coefficient. Receiver operating characteristic (ROC) curves were constructed to determine the sensitivity, specificity, and area under the curve (AUC) of the estimated values compared with the measured values for elevated ICP (ICP ≥ 15 mmHg). Statistical significance was defined as p-value < 0.05.
Results
Participants’ flow chart and numbers analysed
The participants’ flow diagram shown in Figure 2 includes the number of participants evaluated for potential enrolment into the pilot study and the number excluded at this stage either because they did not meet the inclusion criteria or declined to participate. The diagram also displays information regarding the number of patients included in the main analysis, with numbers and reasons for exclusions. Moreover, the authors included an additional step concerning the random allocation of observations to build the machine-learning model.
Recruitment
After ethics approval, 40 patients were enrolled in this single-arm prospective study beginning in January 2020, with completed enrolment in July 2021. All actively participating patients met the inclusion criteria of the study.
Baseline data
Patient demographics and baseline measures are identified in Table 2. The mean patient age was 42 (SD = 16) years old, with 87% male.
Outcomes and estimation
The distribution of the observations for testing and training was skewed toward lower ICP values (Figure 3a), within the 5 to 40 mmHg range of ICP values. More data were available at lower ICP values, especially between 10 and 15 mmHg, because all patients were undergoing active medical treatment to avoid intracranial hypertension as part of their normal medical treatment. The estimated intracranial pressure (nICP) was compared against the invasive ICP reference on the testing group. Figure 3Error! Reference source not found.b shows an uphill pattern from lower to higher ICP values; this indicates a positive relationship between the estimated ICP values and the invasive measurements.
Having a determination coefficient of 0.67 indicates that 67% of the variation in the intracranial pressure is explained by variation in the estimated ICP values. Because the coefficient of determination cannot exceed 100 %, a value of 67 % indicates that the bagging tree matches the actual sample data. Moreover, the estimated and invasive data have a strong correlation of 82.54% (p-value
< 0.05), meaning that an increase in nICP corresponds to an increase in invasive measurement. However, the estimation performance drops for higher ICP values due to lower availability of high ICP data for training the model. Moreover, in order to analyse the agreement between the two methods, a Bland-Altman analysis was performed, comparing the differences in mmHg as shown in Figure 3c. The mean difference, or bias, between nICP and the invasive reference, was of -0.22 mmHg difference. Furthermore, the limits of agreement (Z=1.96) indicate that 95% of the differences between the two monitoring methods are within the range of ±3.8 mmHg. Additionally, the variability of the bias is consistent across the graph, and in accordance with the trend line, the estimated values at low ICP values (<14 mmHg) are higher than the gold standard, while at high ICP (>14 mmHg) are lower. Although an accurate ICP value is useful in the clinical management of TBI patients, the model diagnostic capability to detect hypertensive events could be of interest in neurocritical care. Therefore, Figure 3d presents the ROC curve that evaluated the model performance considering the limitation of an unbalance data distribution within the ICP intervals. Therefore, a lower threshold (15 mmHg) than the one recommended by the clinical guidelines (20 mmHg), was defined. The estimated nICP showed an area under the curve (AUC) of 0.92, a sensitivity of 79.7% and a specificity of 89%.
In addition to the results on the whole dataset, the model was tested individually per patient. Table 3 summarises the mean ± standard deviation and the confidence interval (95%) of the different accuracy parameters evaluated. From this individual analysis, three patients were not included as they did not have more than 100 observations. These results show the variability of the model performance across the patients, presenting reasonable limits of agreement of approximately ± 3.5 mmHg, a mean RMSE of 2.79 ± 0.93 mmHg, and correlation coefficients within 77% and 83% in 95% of the patients.
Figure 3e shows a continuous estimate of ICP (black line) on top of the testing group’s invasive reference (red). The amount of data per ICP interval is noticed in this figure, as well as the trend of the nICP estimation against the gold standard. Similarly, Figure 3f displays the trending of one random patient who experienced increases in intracranial pressure. Again, the red line matches the trend of the invasive measurements with a root mean square error of 3.22 mmHg.
Ancillary analyses
The quality of the cerebral PPG signals is key for extracting reliable features from the PPG waveform. Therefore, the signal processing methods included a denoising algorithm with the aim of removing anomalies from the PPG-NIRS signals, as shown in Figure 4. Data length before and after denoising was reduced by 40.18% from 1425.67 hours before denoising to 852.83 hours. However, the Signal Noise Ratio (SNR) increased after the application of the denoising algorithm by 35%.
Harms
The nICP sensor was attached to the forehead of TBI patients in neurocritical care for up to 48 hours. Near-infrared light was continuously shone into the tissue with a controlled current of up to 30mA. This non-invasive tool did not lead to unintended consequences, harms or effects. Mild marks on the skin were observed after sensor removal, yet no skin lesions or burns were reported. The skin marks were thought to be related to the probe being in place for 48 hours and the marks rapidly disappeared leaving no permanent marks in any participant.
Discussion
This paper reports a novel optical device’s first clinical pilot study to measure ICP non-invasively using PPG-NIRS waveform analysis. It has several advantages over previously described systems, including TCD and ONSD measurements, in that it is truly non-invasive and is neither operator nor patient dependent.
Interpretation
The correlation between the invasive and non-invasive instruments observed in this work showed for the first time that the non-invasive ICP monitor effectively followed the trends in ICP changes during the protocol. This provided reassurance that the features extracted from the nICP signals correlate with changes in ICP as initially hypothesised. The estimated nICP also showed good accuracy and good agreement (i.e. RMSE = 3.3 mmHg and limits of agreement of ± 3.8 mmHg) in estimating absolute values of ICP when compared to the reference ICP. Furthermore, the accuracy of different invasive ICP sensors has been reported to be in the range of 0.7-2.3 mmHg41, whereas the expected accuracy is ± 2 mmHg for ICP<20 mmHg or ±10% for ICP>20 mmHg42. For these reasons, the accuracy between the nICP monitor and the reference invasive ICP sensor observed in this investigation aligned with the expected error for a new ICP measuring technique.
Additionally, this work compares favourably to other non-invasive ICP monitors. For the non-invasive estimation of ICP, a number of TCD-based techniques have been employed. By evaluating a large variety of TCD-based techniques, Cardim and colleagues reported confidence ranges ranging from 4.2 to 59.6 mmHg, with an average confidence interval of about 12 mmHg20. Similarly, Bellner et al. looked for the correlation of the TCD pulsatility index with ICP, which led to an accuracy of ± 4 mmHg43, yet other studies could not reproduce these results. Kim et al. employed a semi-supervised algorithm to categorise intracranial hypertension by analysing the morphology of the TCD waveform, but this approach could not provide a continuous ICP estimate44.
Other recently developed non-invasive techniques either analyse acoustic waves that pass through the skull and provide an ICP estimate accuracy of ±6.8 mmHg 45 or measure with ultrasound or other imaging techniques the optic nerve sheath diameter to detect elevated ICP 46. Most of the ICP monitoring systems addressed in the review by Rosenberg et al. employed CT or ultrasound to assess the diameter of the optical nerve sheath. 47. However, they only make a binary decision between elevated and normal ICP, while the nICP sensor reported in this study has the additional advantage of allowing continuous estimations of ICP instead of binary decisions.
Finally, Ruesch et al. have been working with NIRS and Diffuse Correlation Spectroscopy (DCS) technologies to estimate ICP non-invasively. The authors present a recompilation of studies where ICP changes were induced incrementally (from about 3-10 mmHg up to 40 mmHg in steps of 10 mmHg) through fluid infusion in non-human primates (n=5 to 8)48–50. These studies utilised the cardiac pulse acquired by DCS in a similar way to how this pilot trial used cerebral PPG signals. Interestingly, as explained by Themelis et al., flow and volume are closely correlated51. Therefore it is not surprising that morphological features from the flow pulsations change when ICP increases, as the volume signals from NIRS-PPGs do. Nonetheless, Ruesh’s approach relies on incorporating ECG measurements for pulse detection in the DCS signals, which is not needed by the PPG method as these signals are synchronised to the heart cycle. Moreover, Ruesh’s method is not entirely non-invasive, as it utilises MAP measurements from an arteria line to estimate ICP48–50.
Limitations
Assuming that altered cerebral vessels’ geometry brought on by high pressure is the cause of morphological waveform alterations, it is essential to recall that other mechanisms, such as cerebral autoregulation, can also alter the vasomotor tone. The authors accept that cerebral autoregulation is frequently disrupted in patients with intracranial hypertension however this study was not designed to detect this.
An accuracy of ± 3.8 mmHg in the random allocation method may be unrepresentatively good. This is because some data from all subjects are included in the training set. Future work would allow the analysis of additional models and training methods.
In addition, the regression model’s accuracy depends on the training data set distribution. An increasing error with increasing ICP is an effect of fewer training data points at larger ICP values, which is a limitation of the study as most patients are medicated to avoid intracranial hypertension, however this may not be as clinically important as detecting rises in ICP approaching thresholds of 20 or 25 mmHg (the usual cut off for increasing intervention) since anything above this level will be considered for treatment regardless of how far it is above the threshold.
The results presented in this research were analysed off-line however, near real-time waveform analysis would be possible by deploying the algorithms into an embedded system. Future clinical research on an embedded system could provide a performance evaluation of the model in real-time.
Conclusions
In conclusion, this study has introduced and demonstrated a new and potentially transformative method and an algorithm to estimate ICP in a continuous and truly non-invasive manner which is not co-dependent on any other measurements with cerebral PPG-NIRS. In particular, a bagging tree fed with morphological features from cerebral vessel pulsations that are measured optically by NIRS to estimate ICP values coming from an invasive ICP sensor. The method was demonstrated on adults with TBI.
With further improvements and optimisations of the technology and more validations studies the nICP sensor might be suitable for other conditions including, hydrocephalus, meningitis, and stroke patients, among others. The ability of pre-hospital physicians to initiate ICP-directed neuro-critical care within the “golden hour” (and often several hours before the patient reaches the hospital) may significantly improve outcomes in trauma patients. It may additionally open new scenarios for patients from low- and middle-income countries where the majority of the world’s population has no access to neurosurgery: a low-cost, non-invasive monitor may allow ICP-directed therapy in trauma that could be implemented in any hospital.
Data Availability
All data produced in the present study are available upon reasonable request to the authors
Ending section
Authorship contribution
M.R: Methodology (supporting); data acquisition (equal); software (equal); formal analysis (lead); investigation (equal); writing – original draft (lead). T.A.Y: Conceptualization (lead); methodology (lead); Resources (lead); software (equal); data acquisition (equal); investigation (equal). C.U: Conceptualization (supporting); Methodology (supporting); project administration (supporting); data acquisition (equal); validation (equal); writing – review and editing (equal). P.K: Conceptualization (supporting); validation (equal); project administration (equal); supervision (lead); writing – review and editing (equal).
Conflicts of Interest Statement
The authors declare no conflicts of interest.
Funding Statement
The project is funded by the National Institute for Health Research (NIHR) [Invention for Innovation (i4i) Product Development (Grant Reference Number II-LA-0216-20005)]. The views expressed are those of the authors and not necessarily those of the NIHR or the Department of Health and Social Care.
Acknowledgements
The authors would like to thank Dr Justin P Phillips for his valuable suggestions on applying cerebral PPG signals for the non-invasive estimation of ICP. Similarly, they want to thank George Bradley for his contributions in signal processing.