Abstract
Background Multi-steroid profiling is a powerful analytical tool that simultaneously quantifies steroids from different biosynthetic pathways. Here we present an ultra-high performance liquid chromatography-tandem mass spectrometry (uPLC-MS/MS) assay for the profiling of 25 steroids using post-column infusion of ammonium fluoride.
Methods Following liquid-liquid extraction, steroids were chromatographically separated over 5 minutes using a Phenomenex Luna Omega C18 column and a water (0.1 % formic acid) methanol gradient. Quantification was performed on a Waters Acquity uPLC and Xevo® TQ-XS mass spectrometer. Ammonium fluoride (6 mmol/L, post-column infusion) and formic acid (0.1 % (vol/vol), mobile phase additive) were compared as additives to aid ionisation.
Results Post-column infusion (PCI) of ammonium fluoride (NH4F) enhanced ionisation in a steroid structure-dependent fashion compared to formic acid (122-140% for 3βOH-Δ5 steroids and 477-1274% for 3-keto-Δ4 steroids). Therefore, we fully analytically validated PCI with NH4F. Lower limits of quantification ranged from 0.28 to 3.42 nmol/L; 23 of 25 analytes were quantifiable with acceptable accuracy (bias range −14% to 11.9%). Average recovery ranged from 91.6% to 113.6% and average matrix effects from −29.9% to 19.9%. Imprecision ranged from 2.3% to 23.9% for all analytes and was <15% for 18/25 analytes. The serum multi-steroid profile of 10 healthy men and 10 healthy women was measured.
Conclusions uPLC-MS/MS with post-column infusion of ammonium fluoride enables comprehensive multi-steroid profiling through enhanced ionisation particularly benefiting the detection of 3-keto-Δ4 steroids.
Highlights
This multi-steroid profiling assay quantifies 25 steroids in 5.5 minutes
Post-column infusion of NH4F enhances the ionisation of 3-keto-Δ4 steroids
The assay simultaneously quantifies steroids from several biosynthetic pathways
We present analytical data validated for serum steroid profiling
Competing Interest Statement
The authors have declared no competing interest.
Funding Statement
This study was funded by Wellcome Trust (Investigator Award WT209492/Z/17/Z, to W.A.) and Academy of Medical Sciences UK (Newton Advanced Fellowship NAF004\1002, to K.H.S).
Author Declarations
I confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.
Yes
The details of the IRB/oversight body that provided approval or exemption for the research described are given below:
Science, Technology, Engineering and Mathematics Ethical Review Committee of the University of Birmingham, gave ethical approval for this work ERN_17-0494
I confirm that all necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived, and that any patient/participant/sample identifiers included were not known to anyone (e.g., hospital staff, patients or participants themselves) outside the research group so cannot be used to identify individuals.
Yes
I understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).
Yes
I have followed all appropriate research reporting guidelines and uploaded the relevant EQUATOR Network research reporting checklist(s) and other pertinent material as supplementary files, if applicable.
Yes
Data Availability
All data produced in the present study are available upon reasonable request to the authors