Abstract
Magnetoencephalography (MEG) is an established method to investigate epilepsy. Current MEG systems house hundreds of cryogenic sensors in a rigid, one-size-fits-all helmet, which results in several limitations, particularly in children. On-scalp MEG based on optically-pumped magnetometers (OPMs) may alleviate these limitations.
We report on five children (5–11 years old) with self-limited focal (n=3) or structural (n=2) epilepsy who underwent cryogenic (102 magnetometers) and on-scalp (32 OPMs) MEG. We compared the two modalities for the detection and localization of interictal epileptiform discharges (IEDs).
We identified IEDs in all children with comparable sensor topographies for both MEG devices. IED amplitudes were 2.3-4.8 times higher with on-scalp MEG and signal-to-noise ratio (SNR) was also 27-60% higher with on-scalp MEG in all but one patient with large head movement artifacts. The neural source of averaged IEDs was located at about 5 mm (n=3) or higher (8.3 mm, n=1; 15.6 mm, n=1) between on-scalp and cryogenic MEG.
Despite limited number of sensors and scalp coverage, on-scalp MEG detects IEDs in epileptic children with higher SNR than cryogenic MEG. This technology, which is in constant development, should become a reference in the diagnostic workup of epilepsy and replace cryogenic MEG in the near future.
Competing Interest Statement
Niall Holmes (Scientific Advisor) and Matthew Brookes (Chairman) are involved in CERCA Magnetics Limited (https://www.cercamagnetics.com/). The other authors have no conflict of interest.
Funding Statement
Odile Feys is supported by a research grant from the Fonds pour la Formation a la Recherche dans l'Industrie et l'Agriculture (FRIA, Fonds de la Recherche Scientifique (FRS-FNRS), Brussels, Belgium). Pierre Corvilain is supported by a research grant from the Fonds Erasme (Research Convention Alzheimer, Brussels, Belgium). Xavier De Tiege is Postdoctorate Clinical Master Specialist at the FRS-FNRS. This study has been supported by the Fonds Erasme (Research Conventions Les Voies du Savoir & Projet de recherche clinique a des techniques medicales emergentes 2020) and the FRS-FNRS (Equipment credit : U.N013.21F). The OPM- and SQUID-MEG projects at the CUB Hopital Erasme are financially supported by the Fonds Erasme.
Author Declarations
I confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.
Yes
The details of the IRB/oversight body that provided approval or exemption for the research described are given below:
Ethics Committee of the CUB Hopital Erasme (Reference: P2019/426, B406201941248)
All necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived.
Yes
I understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).
Yes
I have followed all appropriate research reporting guidelines and uploaded the relevant EQUATOR Network research reporting checklist(s) and other pertinent material as supplementary files, if applicable.
Yes
Data Availability
Data are available upon reasonable request to the corresponding author and after approval of institutional (i.e., CUB Hopital Erasme & Universite libre de Bruxelles) authorities.
Abbreviations
- EZ
- Epileptogenic zone
- IED
- Interictal epileptiform discharge
- MEG
- Magnetoencephalography
- MSR
- Magnetic shielded room
- OPM
- Optically-pumped magnetometer
- OPM-MEG
- Optically-pumped magnetometer based magnetoencephalography
- SL-ECTS
- Self-limited epilepsy with centro-temporal spikes
- SNR
- Signal-to-noise ratio
- SQUID
- Superconducting quantum interference device
- SWI
- Spike-wave index