Abstract
Background and Purpose Head motion causes image degradation in brain MRI examinations, negatively impacting image quality, especially in pediatric populations. Here, we used a retrospective motion correction technique in children and assessed image quality improvement for 3D MRI acquisitions.
Material and Methods We prospectively acquired brain MRI at 3T using 3D sequences, T1-weighted MPRAGE, T2-weighted Turbo Spin Echo and FLAIR, in 32 unsedated children, including 7 with epilepsy (age range 2–18 years). We implemented a novel motion correction technique: Distributed and Incoherent Sample Orders for Reconstruction Deblurring using Encoding Redundancy (DISORDER). For each subject and modality, we obtained 3 reconstructions: as acquired (Aq), after DISORDER motion correction (Di), and Di with additional outlier rejection (DiOut).
We analyzed 288 images quantitatively, measuring 2 objective no-reference image quality metrics: Gradient Entropy (GE) and MPRAGE White Matter Homogeneity (WM-H). As a qualitative metric, we presented blinded and randomized images to 2 expert neuroradiologists who scored them for clinical readability.
Results Both image quality metrics improved after motion correction for all modalities and improvement correlated with the amount of intrascan motion. Neuroradiologists also considered the motion corrected images as of higher quality (Wilcoxon’s z = −3.164 MPRAGE, z = −2.066 TSE, z = −2.645 FLAIR, for all p < 0.05).
Conclusions Retrospective image motion correction with DISORDER increased image quality both from an objective and qualitative perspective. In 75% of sessions, at least one sequence was improved by this approach, indicating the benefit of this technique in unsedated children for both clinical and research environments.
Competing Interest Statement
The authors have declared no competing interest.
Clinical Trial
CPMS40005
Funding Statement
This research received funding from a Sir Henry Dale Fellowship jointly funded by the Wellcome Trust and the Royal Society (Grant Number 206675/Z/17/Z). The study was also supported in part by the Wellcome Engineering and Physical Sciences Research Council Centre for Medical Engineering at King s College London (grant WT 203148/Z/16/Z), the Medical Research Council (UK) (grants MR/K006355/1 and MR/LO11530/1) and Medical Research Council Centre for Neurodevelopmental Disorders, King s College London (MR/N026063/1). Infrastructure support was provided by the National Institute for Health Research Mental Health Biomedical Research Centre at South London, Maudsley NHS Foundation Trust, King s College London, the National Institute for Health Research Mental Health Biomedical Research Centre at Guys, and St Thomas Hospitals NHS Foundation Trust.
Author Declarations
I confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.
Yes
The details of the IRB/oversight body that provided approval or exemption for the research described are given below:
Approval for this study was obtained by the Health Research Authority (HRA) and Health and Care Research Wales (HCRW), ethics ref 18/LO/1766.
All necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived.
Yes
I understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).
Yes
I have followed all appropriate research reporting guidelines and uploaded the relevant EQUATOR Network research reporting checklist(s) and other pertinent material as supplementary files, if applicable.
Yes
Abbreviations
- DISORDER
- Distributed and Incoherent Sample Orders for Reconstruction Deblurring using Encoding Redundancy
- MPRAGE
- Magnetization Prepared Rapid Acquisition Gradient Echo
- FLAIR
- Fluid Attenuated Inversion Recovery
- TSE
- Turbo Spin-Echo
- GE
- Gradient Entropy
- WM-H
- White Matter Homogeneity
The Chan Zuckerberg Initiative, Cold Spring Harbor Laboratory, the Sergey Brin Family Foundation, California Institute of Technology, Centre National de la Recherche Scientifique, Fred Hutchinson Cancer Center, Imperial College London, Massachusetts Institute of Technology, Stanford University, University of Washington, and Vrije Universiteit Amsterdam.