Abstract
Background Artificial neural networks have shown promising performance in automatic segmentation of cardiac magnetic resonance imaging. However, initial training of such networks requires large amounts of annotated data and generalization to different vendors, field strengths, sequence parameters, and pathologies is often limited. Transfer learning has been proposed to address this challenge, but specific recommendations on the type and amount of data required is lacking. In this study we aim to assess data requirements for transfer learning to cardiac 7T in humans where the segmentation task can be challenging. In addition, we provide guidelines, tools, and annotated data to enable transfer learning approaches of other researchers and clinicians.
Methods A publicly available model for bi-ventricular segmentation is used to annotate a publicly available data set. This labelled data set is subsequently used to train a neural network for segmentation of left ventricular and myocardial contours in cardiac cine MRI. The network is used as starting point for transfer learning to the segmentation task on 7T cine data of healthy volunteers (n=22, 7873 images). Structured and random data subsets of different sizes were used to systematically assess data requirements for successful transfer learning.
Results Inconsistencies in the publically available data set were corrected, labels created, and a neural network trained. On 7T cardiac cine images the initial model achieved DICELV=0.835 and DICEMY=0.670. Transfer learning using 7T cine data and ImageNet weight initialization significantly (p<10−3) improved model performance to DICELV=0.900 and DICEMY=0.791. Using only end-systolic and end-diastolic images reduced training data by 90%, with no negative impact on segmentation performance (DICELV=0.908, DICEMY=0.805).
Conclusions This work demonstrates the benefits of transfer learning for cardiac cine image segmentation on a quantitative basis. We also make data, models and code publicly available, while providing practical guidelines for researchers planning transfer learning projects in cardiac MRI.
Introduction
Image segmentation, which is of great interest in cardiac magnetic resonance imaging is applied to partition acquired images into functionally meaningful regions, allowing the extraction of quantitative static measures such as myocardial mass, left ventricle (LV) volume, right ventricle (RV) volume, and wall thickness, as well as dynamic measures such as wall motion and the ejection fraction (EF). Cardiac cine MRI is the accepted gold standard for this assessment of cardiac function1 and anatomy and is therefore of paramount clinical importance2,3. Proper segmentation of such data sets is a tedious and time-consuming process that has increasingly been tackled using various deep learning approaches4-7.
Artificial neural networks have been shown to outperform other methods on several high profile image analysis benchmarks and, thus, so-called deep learning models have become state-of-the-art for a wide variety of computer vision tasks. Multiple factors like the wide application area of deep learning, available compute power, and increasing investments as well as user-friendly open source software have enabled a rapid development of the field of artificial intelligence. This led to ever increasing applications in medical imaging such as MRI8 where tasks nowadays range from data acquisition and image reconstruction9-11, image restoration12, 13, to image registration14, 15, segmentation16-19 as well as classification20, 21 and outcome prediction22,23.
There is consensus in the field that the limited availability of labelled or annotated data due to data access, privacy issues, missing data harmonization, and data protection is one of the main obstacles for future clinical applications of deep neural networks17, 19, 24. While some resources like the UK Biobank25 already exist to address this issue, the high quality standards and the amount of work required to organize and maintain such a resource makes data access expensive. In addition, such data may already exceed the quality that is available in clinical routine cardiac MRI. This leads to neural networks, which perform very well for a very specific task within a confined data space, where training and testing data share the same distribution. However, these networks usually lack generalization capabilities. While methods such as data augmentation, transfer learning, weakly-, self-supervised, and unsupervised learning have been applied to overcome the issue of small datasets in research, it is unclear how much data is really required in order to create a well-generalizing network or to apply transfer learning.
In this work, we aim to enable researchers and clinicians in cardiology to apply deep learning-based segmentation models in their respective research by providing guidelines and easily accessible tools as well as annotated data for transfer learning. We create labels for a public data set, the Data Science Bowl Cardiac Challenge Data26 (further referred to as Kaggle data set) which, at this point, does not have segmentation labels. We further create a base network for LV segmentation using these labels and evaluate its performance on 7T human cine data. In addition, we assess if transfer learning improves model performance for the 7T segmentation task and analyze how much and which data is required. The framework provided in this study in combination with access to scripts and the data used, will enable researchers to reproduce our results and apply deep learning based segmentation in their respective field.
Methods
The Kaggle Data Set
As mentioned above, cardiac MRI is the gold standard for the assessment of cardiac function, a key indicator of cardiac disease. The 2015 Data Science Bowl challenged participants to create an algorithm for automatic assessment of end-systolic and end-diastolic volumes (ESV and EDV) and thus, ejection fraction, based on cardiac cine MRI. The data set consists of a training, a validation, and a test set and once the challenge has ended, all sets and their corresponding volume information (end-systolic and end-diastolic) was made available for research and academic pursuits, leading to a total of 1140 “annotated” cardiac MRI examinations of normal and abnormal cardiac function. Images are in DICOM format resolving up to 30 phases of the cardiac cycle. While we will focus on short axis images in this study, the Kaggle data set also contains alternative views. Examinations were done on 1.5 T and 3.0 T systems (Siemens Magnetom Aera and Skyra, Siemens Healthineers, Erlangen, Germany) with applications of both FLASH and TrueFISP sequences. An overview of the complete data set and its variation in patient data and sequence parameters is given in Table 1.
Data Composition and measurement parameters of the Kaggle data
Data Curation
The complete data set is a compilation of real, clinical data from several sites and as such, subject to inconsistencies within individual examinations. Those can be a combination of:
▪ missing time points (e.g. all slices with 30 cardiac phases and one with 25)
▪ inconsistent slice spacing
▪ inconsistent image dimension
▪ repeat measurements (identical slice location)
▪ scaled images
▪ image rotations
Prior to the application of the published segmentation network of Bai et al.4 we performed data curation, correcting inconsistencies in all but 8 examinations. More detailed information and curated data can be found in the online repository (https://github.com/chfc-cmi/cmr-seg-tl, https://doi.org/10.5281/zenodo.3876351).
Creating Labels
Once the data was corrected for inconsistencies we ran the Python based segmentation model of Bai et al. for the complete data set, generating RV, LV, and blood pool labels as well as LV ESV and EDV volumes. ESV and EDV values were then compared to the ground truth values provided by Kaggle in order to determine the accuracy of the network prediction. Based on this comparison we created confidence sets where the predicted values were in the range of ±5% (p5), ±10% (p10), and ±15% (p15) of the true value. Respectively, these sets contained 175, 520, and 763 examinations and 54540, 162480, and 238350 images. All scores (label versus ground truth) for ESV and EDV values are listed in the online repository.
Hardware
In order to deal with the extensive computation demands we used a custom workstation and a high performance cluster, both with graphical processing units. Details are given in the online repository.
Framework - Deep neural network
All implementations were realized using Pytorch27 and fastai28 V1. Training of neural networks (U-Net29 architectures with varying backbones: Restnet3430, ResNet5031, and VGG1632) was performed using fastai’s implementation of the one cycle policy33 with adjusted learning rates (lr) and the confidence sets p5, p10, and p15.
Parameter Search
During the parameter search, we evaluated the influence of different training parameters on the efficacy of the trained model. Training speeds of the varying models and their architectures is given in the online repository. Training with a weight-decay of 0.02 and a batch-size of 32 was done for 30 epochs with frozen weights (lr = 1e-4) and another 30 epochs with unfrozen weights (lr = 1e-5). Details regarding frozen and unfrozen weights are provided in the online repository. The smallest training set (p5) was used initially, image size was 256×256, and moderate data augmentation transforms (s1: flip [none], rotation [20°], lighting [0.4], zoom [1.2], padding [zeros]) were applied.
In order to avoid an extensive parameter grid search, we assessed parameter dependent performance changes in incremental steps. After each step, we determined the best-performing model using EF predictions and introduced subsequent parameter variations on this respective model.
In the first step we evaluated the influence of the architecture (VGG16, ResNet34, ResNet50) compared to the fully convolutional Network by Bai et al4. trained on UKBB data (further referred to as UKBB model). Due to memory limitations, we had to reduce the batch size for training of the VGG16 and the ResNet50 models.
In the second step, we assessed variations in the loss function such as cross-entropy (default), generalized DICE34, and focal loss. In the third and last step we evaluated the influence of the number of training images using the confidence sets p5, p10, and p15.
We assessed the influence of training data resolution, training a model with lower input resolution (128×128, r34_CE_p5_128). Details are provided in the online repository.
Data Augmentation
Since transfer learning applications assessed in this study are based on 7T data we expect somewhat different image contrast and artefacts compared to conventional, clinical datasets. In addition, we intended to account for the heterogeneous training data, which led to the following set of augmentations for the initial networks (s1: flip [none], rotation [20°], lighting [0.4], zoom [1.2], padding [zeros]). Further, we aimed to introduce some robustness to forms of data variations, such as 90°-rotations and flips (left-right) using more extensive data augmentation (s2: flip [Left-Right], rotation [90°], lighting [0.4], zoom [1.2], padding [zeros]). In order to test the efficacy of these transforms we trained a new model (r34_CE_p5_s2) and compared EF predictions on a dataset including rotated and flipped images retained during the data curation process.
Transfer Learning
All assessments regarding transfer learning to 7T data are done using model: r34_CE_p5_s2. As initial point of comparison we used the UKBB model to create labels for 7T data, in order to assess generalization capability of a model, which was trained on a very homogeneous data set (UKBB).
Following approval of the local ethics committee (7/17-SC), n=22 (14 female, 8 male) were examined using a 7T whole body MRI system (Siemens MAGNETOM Terra, Erlangen, Germany) and a 1TX/16RX thorax coil (MRI Tools, Berlin, Germany)35. Written informed consent was obtained prior to all measurements. Patient age was 22-53 years, body weight 52-95 kg, and height: 151-185cm. For triggering, both the integrated ECG and an external acoustic triggering system (MRI Tools, Berlin, Germany) were used in order to synchronize measurements with the heartbeat, choosing whichever method provided a more stable trigger signal during the examination. Images were obtained using a cardiovascular (CV) GRE cine-sequence. Sequence parameters were: TE = 3.57ms, FOV = 340 mm × 320 mm, interpolated voxel size = 0.66 × 0.66 × 6 mm, GRAPPA acceleration factors: R = 2 and R = 3. Depending on the heart rate 6-11 segments and 20-35 cardiac phases were measured using retrospective gating. Short axis CINE stacks for volumetric evaluation varied in the number of slices (14-17) and multiple breath-holds (∼13s) were necessary to acquire the whole stack. Images were assigned into training, validation and test sets (14, 5, 3 subjects and 5076, 1842, 955 images, respectively). All images were manually segmented by an expert radiologist (TR). Three data sets of the test set were additionally segmented by an expert cardiologist (WS), in order to obtain an estimate of interobserver-variability.
Starting Point for Model Training - 7T Human
To assess the efficacy of transfer learning for LV segmentation based on clinical 1.5T and 3T data and experimental (human) 7T data, we compare models with varying degrees of training and transfer learning. Using a U-Net architecture with a ResNet34 backbone (r34_CE_p5_s2), we generated the following three models:
initialization with random weights (R)
initialization with ImageNet-weights – transfer learning 1 (TL)
Model 2, pre-trained on Kaggle data - transfer learning 2 (TL2)
All models were used to generate predictions for the 7T test set.
Model performance was always evaluated using the Soerensen-DICE36 coefficient between predictions and respective ground truth labels.
Data Requirements for Model Training - 7T Human
To assess how much and what data is required for convergence of a model we trained all models (R, TL, TL2) with subsets of the training data. These subsets were created in two ways:
Complete subject data (all slices and all phases) from 14, 7, 3, 1 subjects (5076, 2626, 1001, 306 images, respectively); Partial subject data (only end-systolic and end-diastolic images) from all subjects (448 images)
Shuffle all images once, create a list of images (1-5076), and generate subsets corresponding to the respective image numbers from subset 1, always starting the count with image #1
When training with subsets, the model is exposed to a smaller number of images in every epoch. We therefore increased the number of epochs for the subsets to correct for this effect.
Results
Framework - Deep neural network
Parameter Search
Results of the parameter search are illustrated in Figure 1, showing the absolute distance between the EF predictions based on model segmentation and ground truth data provided by Kaggle. Overall the impact of parameter variation on model performance was small (3.64-4.06% mean distance to ground truth EF).
Plots show the absolute distance between the EF prediction based on model segmentation and ground truth data provided by Kaggle. The range of the y-axis is restricted for better comparability, dashed lines indicates lowest median. Model performance with A: Architectures (r34: ResNet34, r50: ResNet50, VGG16: v16, UKBB). B: Loss functions (Cross-entropy: CE, DICE, focal loss). C: Confidence sets (p5: 5%, p10: 10%, p15: 15%).
In a first approach to interpret these results, we compared varying architectures, such as ResNet34, ResNet50, and VGG16 with the UKBB model (Figure 1A). All models led to lower mean and median distance values compared to the UKBB model (table 2). The lowest median distance values were found using a ResNet50 (2.79%), while the lowest mean distance values were found using a ResNet34 (3.64%). Differences in the absolute distance between the models (r34, r50) were rather small (Δ0.08%), however. Considering computational demand, we selected the ResNet34.
Summary statistics of absolute deviation of predicted and true EF in % for the parameter search. Sorted from lowest to highest mean value. Models are named by architecture (ResNet34: r34, ResNet50: r50, VGG16: v16), loss function (cross entropy: CE, focal, DICE), confidence set (p5, p10, p15), and data augmentation (s1, s2).
In the next step of the parameter search we evaluated model performance using varying loss functions, namely cross-entropy, generalized DICE, and focal loss (Figure 1B). Using the generalized DICE score led to the highest mean (3.93%) and median (3.07%) distance values. Median distance values were similar for cross-entropy and focal loss (2.87% vs 2.86%), while the mean distance value was lowest using cross-entropy (3.64%).
We thus selected cross-entropy for the next step of the parameter search, where we evaluated model performance using varying confidence sets: 5%, 10%, 15% (Figure 1C). Using the various confidence sets only slightly affected median distance values (2.87%, 2.89%, 2.91%). Based on EF predictions the model: r34_CE_p5_s1 performed best achieving a mean distance value of 3.64%.
Data Augmentation
Figure 2 shows the performance of our models on the image set containing rotated images, plus the performance of an additional model where data augmentation allowed left-right flips, as well as rotations of up to 90°. Median and mean absolute distance values were lowest (3.06%, 4.08%) using the model with extended data augmentation.
Plots show the absolute distance between the EF predictions based on model segmentation and ground truth data provided by Kaggle for all models of the parameter search, plus one model trained with extended data augmentation (s2). Models are named by architecture (ResNet34: r34, ResNet50: r50, VGG16: v16), loss function (cross entropy: CE, focal, DICE), confidence set (p5, p10, p15), and data augmentation (s1, s2).
Transfer Learning
Exemplary cine images from the Kaggle and the 7T cine data set as well as respective data augmentation are shown in Figure 3A and 3B. While the Kaggle data set includes images with varying field of views and resolution, the 7T data is consistent.
Random selection of five images (top) with five data augmentation examples (bottom) for the first image of the random selection. A: Kaggle data. B: 7T human cine data.
Figure 4 presents the inter-observer variability as difference of LV volume within each image in ml for all slices and phases of the 7T human cine test set (n=3). The slice count starts with 0 at the most apical slice and moves towards the most basal slice with increasing slice number. Overall the two experts achieved DICELV = 0.94 and DICEMY = 0.81 and deviations in LV volume of individual images were lower than ±5 ml in all but one image (set 3, slice 12, phase 4). Compared to expert 1, who labelled our training data, and expert 2 the AI model achieved DICELV = 0.90, DICEMY = 0.79 as well as DICELV = 0.91, DICEMY = 0.81. Deviations in LV volume of individual images were smaller than 5 ml in >95% of the cases. Exemplary predictions of the AI and deviations to expert 1 are shown in Figure 5. All apical slices labelled by the AI were in excellent agreement with that of our experts. The largest deviations between AI and both experts was found for the very basal slice where myocardial tissue moves in and out of plane throughout the cardiac cycle.
The slice count starts with 0 at the most apical slice and moves towards the most basal slice with increasing slice number. Top: Inter-observer variability of the two experts. Middle: Inter-observer variability expert 1 (labelled training data as well) versus AI. Bottom: Inter-observer variability expert 2 versus AI.
Examples of mid-cavity segmentation results with high (top), intermediate (middle) and low (bottom) DICE scores. Images (left), with predicted classes (middle, background: purple, LV: blue, MY: yellow) and differences to the ground truth (right, LV-error: blue, MY-error: yellow)
Starting Point for Model Training
Results of model training using varying degrees of transfer learning are displayed in Figure 6. Plotted are the DICE scores for the left ventricle and the myocardium in dependence of the number of images seen during training, showing performance and overall convergence for the three models analyzed. All curves have been smoothed to increase interpretability. Respective plots of the raw data are shown in the online repository.
Varying degrees of transfer learning (R: ResNet34 initialized with random weights and trained using 7T cine images, TL: ResNet34 initialized with ImageNet weights and trained using 7T cine images, TL2: ResNet34 initialized with ImageNet weights, pre-trained on the 1.5T and 3T Kaggle cine images and re-trained on 7T cine images) are shown for the two subsets 1 (line): subset of whole volunteers (full=14, 7, 3, 1), 2 (dotted line): subset of random images with image numbers corresponding to first subset. In addition, there is one model (“esed”) trained using only end-systolic and end-diastolic images from all volunteers and a corresponding model trained with a number of random images equivalent to the “esed”-set.
Starting with the full data set, there are clear differences in starting points (DICE after first epoch), convergence speed, and peak performance (highest performance reached) for the three models.
R: Random weight initialization followed by training using 7T data led to the:
lowest starting points with DICELV ∼ 0.57 and DICEMY ∼ 0.25
lowest convergence speed, reaching plateau values after 100.000 images
lowest peak performance with DICELV ∼ 0.89 and DICEMY ∼0.77
TL: ImageNet weight initialization followed by training using 7T data led to the:
starting points of DICELV ∼ 0.77 and DICEMY ∼ 0.51
higher convergence speed, reaching plateau values after 50.000-60.000 images
higher peak performance with DICELV: 0.91 and DICEMY: 0.79
TL2: ImageNet weight initialization, pre-trained (Kaggle data), re-trained 7T data led to the:
highest starting points with DICELV ∼ 0.90 and DICEMY ∼ 0.78
higher convergence speed, reaching plateau values after 40.000-50.000 images
higher peak performance with DICELV: 0.92 and DICEMY: 0.81
Data Requirements for Model Training
Results of model training using varying degrees of transfer learning and a smaller amount of training data are displayed in Figure 6 as well. The full training data set consists of 14 volunteers, while the subsets consist of 7, 3, and 1 volunteer. For the most part, curves follow the trend described for the full data set, while each reduction in volunteers led to lower starting points. Peak performances remain similar with a reduction to 7 volunteers, but drop using subset n3, in particular for models R and TL. Only for a very small number of training images (n1) peak performances are higher for model R (DICELV: 0.86; DICEMY: 0.72) compared to TL (DICELV: 0.83; DICEMY: 0.70).
For small subsets, such as n3 and n1, starting points as well as peak performances of all models is higher using the random selection of training images instead of all images from a set (3/1) of volunteers. The same trend is shown for the set n7 using models R and TL.
Using only end-systolic and end-diastolic images led to similar convergence speed and peak performance regarding DICE scores compared to the full data set (RLV,MY: 0.90, 0.77; TLLV,MY: 0.90, 0.78; TL2LV,MY: 0.92, 0.81 versus RLV,MY: 0.89, 0.77; TL LV,MY: 0.91, 0.79; TL2 LV,MY: 0.92, 0.81). In addition, the selection of end-systolic and end-diastolic images led to increased DICE-scores as starting points, fast convergence, and higher peak performance for all models, when compared to the same number of randomly selected images.
Discussion
In this study, we successfully used a specialized, publicly available model4 to produce labels for a public data set of clinical 1.5 and 3T cardiac cine MRI, enabling access to more annotated data. Based on these labels we created a basic AI model, other researchers can use for their individual segmentation tasks. In addition, we applied transfer learning to segmentation of 7T human cine data, demonstrating that models based on these labels and a moderate amount of new domain data enable state-of-the-art segmentation results.
One of the obstacles to get started in deep learning based segmentation is the large amount of annotated data required to train an initial model. In this study we circumvent this problem by using the public Kaggle data set, to which we provide labels. The quality of these labels was evaluated using the volume information (end-systolic and end-diastolic volumes) included in the original Kaggle data set. Therefore, careful data curation had to be applied to avoid data inconsistencies (slice spacing, changes in image dimensions and image resolution, as well as missing slices) within individual patients. In addition, we found that label quality was connected to image orientation and image resolution. Scores (mean distance between labels and Kaggle “ground truth”), data curation scripts, as well as labels are provided in the online repository, enabling future use in other studies. We want to point out that label quality and accuracy was assessed via comparison to volume information only, with rare exceptions of visual confirmation. Thresholds of 5%, 10%, and 15% (deviation to the “ground truth”) for the subsets used in this study were chosen arbitrarily. With 54540, 162480, and 239350 images respectively, we assumed these three sets to provide the reasonable compromise between label accuracy and label quantity needed to assess data requirements in this specific transfer learning application.
Based on the now annotated data we trained initial segmentation models with varying architectures (ResNet34, ResNet50, VGG16), varying loss functions (cross-entropy, generalized DICE, focal loss), varying training sets (p5, p10, p15). The final model we selected was a ResNet34, using cross-entropy as a loss function, and the p5 set for training with an image resolution of 256×256. While we selected this model based on performance (mean distance to ground truth EF), overall impacts of parameter variations (3.64-4.06% mean distance to ground truth EF) were rather small. Similar to the use in this study, researchers or clinicians can use this model as a starting point for their respective transfer learning applications.
Considering the performance of this model on 7T human cine data (DICELV: 0.84, DICEMY: 0.67), generalization capability appears limited. This is also true for the UKBB model (7T human cine, DICELV: 0.67, DICEMY: 0.52). As the authors4 point out, the UKBB model was “trained on a single data set, the UK Biobank dataset, which is a relatively homogenous dataset” and might therefore “not generalize well to other vendor or sequence datasets”. With respect to the performance on 7T data this just means that, compared to the UKBB dataset, the Kaggle data set contains image patterns and characteristics more similar to the 7T data we acquired. In addition, it emphasizes why improvements in generalization37-39 are needed and why we applied an additional step of transfer learning to 7T data.
Due to differences in training data our initial models based on UKBB labels outperformed the UKBB model on the Kaggle data. While the UKBB model was trained on the homogeneous UKBB data, our models were trained on the heterogeneous Kaggle data itself. In addition, we applied data augmentation with respect to rotations and contrast and used only Kaggle data with the most accurate (top 15%) labels.
While multiple studies4, 5, 26, 40 have demonstrated great image segmentation results for one specific dataset, these models have not been tested on other datasets or initially lack generalization capability. In this study, we show that transfer learning leads to improved model performance. DICE scores achieved on 7T human cine data prior to and after transfer learning were DICELV: 0.84, DICEMY: 0.67 and DICELV: 0.92, DICEMY: 0.81, respectively. This was comparable to human inter-observer variability (DICELV: 0.94 and DICEMY: 0.81) and is within the range of state-of-the-art results, despite the relatively small set of training data19. In addition, inter-observer-variability in EDV (3.5%) and ESV (10.5%) between our model and the expert radiologist are in good agreement with literature reports (EDV: 2.5-5.3%, ESV: 6.8-13.9%)41 based on SSFP CMR imaging.
Typically, segmentation of the left ventricle is done to evaluate ejection fraction, a clinically used parameter. In this study we show that the model based volume prediction on the test set is very accurate for apical, mid-cavity and basal slices, with the exception of the most basal slice, where myocardial tissue moves in and out of plane throughout the cardiac cycle. Since we do not have a “ground-truth” segmentation for the Kaggle data and no information on labelling protocols, we do not know if there is any consistency in the definition of basal slices or the inclusion or exclusion of papillary muscle.
While transfer learning allows models to adapt to similar tasks and new datasets, containing new characteristics and patterns, this step also requires new labels. This aspect is often a limitation, since labelled medical data is difficult to acquire, particularly in areas that require domain-specific knowledge. In addition, the manual labelling process for high quality segmentations itself is often tedious and labor intensive. In this study we show that transfer learning applications (ImageNet weights to Kaggle data to 7T data) for cardiac cine segmentation of human 7T data can provide state-of-the-art results when training with labelled data from 7-14 volunteers (2626 – 5076 images), reaching DICELV: 0.92 and DICEMY: 0.81 as well as accurate EF values. Having labels for three volunteers (1001 images) leads to decent results (DICELV: 0.91 and DICEMY: 0.80). We consider labels for only one volunteer to be insufficient (DICELV: 0.88 and DICEMY: 0.77).
For small training datasets (n≤1001) we show that a random selection of images from multiple volunteers leads to better performance compared to the selection of all images from a smaller number of volunteers (n=3 or n=1, figure 6). Generalization capabilities of a model increase with the amount of variation provided in the training data and thus using data from a multitude of patients or volunteers, where morphology and therefore image content and contrast differ, may be more beneficial than providing the same number of more coherent images from a small number of volunteers. Furthermore we demonstrate that the number of required images can drastically be reduced (from 5076 to 448 images), using labelled data from specific heart phases, end-diastolic and end-systolic, instead of all images. This may be possible, because knowing the two extreme states of contraction the model can deal more easily with intermediate states. Considering that n=448 images (roughly two cardiac EF examinations) enable close to state-of-the-art results for cardiac cine segmentation, data requirements for transfer learning applications in closely related tasks are low. In addition, labels for end-diastolic and end-systolic images are created in routine clinical cardiac examinations and thus easily accessible.
In summary, how much and which kind of data should be included in the transfer learning process should be carefully considered prior to labelling new data. In particular, the notion to provide data patient by patient may result in higher data requirements than necessary. There are various other routine cardiac MR examinations such as T2, T1, LGE, and even T2* that require segmentation38, 39, 42. Transfer learning applications to image segmentation of such varying contrasts may benefit from the amount of annotated data and the framework provided in this study.
With respect to future use of this annotated data we recommend researchers take the following steps:
use the pre-trained model we provide (r34_CE_p5_s2)
re-train with training data from the new domain and tune hyper parameters using validation data from the new domain
evaluate model performance on a test set from the new domain
In this study, we used only the 5-15% of the most accurate kaggle labels to create our base models. Thus, researchers attempting to train their own base network using the labelled Kaggle data should always assess label quality.
There are some limitations connected to the use of the Kaggle dataset. While there are variations in measurement parameters, such as resolution, FOV, matrix size, TE, TR, bandwidth, and slice thickness, most examinations (∼90%) were done at 1.5T. In addition, all data was acquired using Siemens whole body MRI systems. Models trained using this dataset might thus not generalize well to other vendor datasets, requiring transfer learning as demonstrated in this study.
Since no disease-related information is provided in the Kaggle dataset, we have no knowledge which and how many pathological patterns are currently represented in the dataset. In this study we demonstrate that transfer learning to 7T data of healthy human volunteers enables DICE scores of DICELV: 0.92 and DICEMY: 0.81. A clinical application would require a performance assessment or transfer learning for specific cardiac pathologies, both beyond the scope of this cardiology-related methodological work.
Furthermore, the accuracy of the labels we created was assessed based on comparison to provided volume information only and visual confirmation of the contours may be biased, because we do not know if the provided volume information is based on consistent definitions of basal slices or the inclusion or exclusion of papillary muscle. This should be considered when creating models based on this dataset. In general, there is a need for a standard benchmark dataset, where labels are based on standardized protocols and images are representations of diverse clinical phenotypes (diseases, vendors, field strengths, sequences, protocols).
Conclusion
In this study, we provide access to annotated cardiac cine MRI data, and AI models, which can be used as a starting point for transfer learning applications. Using such a base model, we demonstrate that transfer learning from clinical 1.5 and 3T cine data to 7T cine data is feasible with moderate data requirements, enabling future applications to other cardiac MRI examinations such as T2, T1, LGE, and even T2*. Furthermore, we show that not all data has the same value with respect to transfer learning approaches and that careful selection of the training data may drastically reduce data requirements.