Abstract
The potential for acute shortages of ventilators at the peak of the Covid-19 pandemic has raised the possibility of needing to support two patients from a single ventilator. To provide a system for understanding and prototyping designs we have developed a mathematical model of two patients supported by a mechanical ventilator. We propose a standard setup where we simulate the introduction of T-splitters to supply air to two patients and a modified setup where we introduce a variable resistance in each inhalation pathway and one-way valves in each exhalation pathway. Using the standard setup, we demonstrate that ventilating two patients with mismatched lung compliances from a single ventilator will lead to clinically-significant reductions in tidal volume in the patient with the lowest respiratory compliance. Using the modified setup, we demonstrate that it could be possible to achieve the same tidal volumes in two patients with mismatched lung compliances, and we show that the tidal volume of one patient can be manipulated independently of the other. The results indicate that, with appropriate modifications, two patients could be supported from a single ventilator with independent control of tidal volumes.
Competing Interest Statement
The authors have declared no competing interest.
Funding Statement
This work was supported by the Wellcome/EPSRC Centre for Medical Engineering [WT 203148/Z/16/Z].
Author Declarations
All relevant ethical guidelines have been followed; any necessary IRB and/or ethics committee approvals have been obtained and details of the IRB/oversight body are included in the manuscript.
Yes
All necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived.
Yes
I understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).
Yes
I have followed all appropriate research reporting guidelines and uploaded the relevant EQUATOR Network research reporting checklist(s) and other pertinent material as supplementary files, if applicable.
Yes
Data Availability
All code is available on https://github.com/splitvent/splitvent)