Abstract
BACKGROUND Jet nebulizers are commonly used to provide continuous aerosolized medication therapy (CAMT). We observed the function of our CAMT system that utilizes the Aeroneb Solo nebulizer system (Aerogen Ltd, Galway, Ireland). METHODS: An observational study was performed on 2 CAMT systems with 15 Aeroneb nebulizers for each system. CAMT was simulated for 1, 2 and 3 hours. Continuous nebulization was monitored and residual volumes were recorded at the end of each simulation. Our primary endpoint was established as intermittent nebulization observed by nebulizer filling of > 1 ml during CAMT simulation. Secondary endpoint was a residual volume of < 0.1 ml.
RESULTS Out of 30 simulations in two arms, a fluid level was observed to accumulate intermittently in three nebulizers with a residual volume of 0.7 mls in one of these three. This produced a total success rate of 90%, Arm-A 80%, Arm B-100%, for our primary endpoint. Our secondary endpoint was achieved in 29 of the 30 nebulizers for an overall 97% success rate, Arm A-93%, Arm B-100%. CONCLUSION: Our Aerogen Solo CAMT system successfully emitted the set dose with 90% accuracy.
Competing Interest Statement
The authors have declared no competing interest.
Funding Statement
No funding used
Author Declarations
All relevant ethical guidelines have been followed; any necessary IRB and/or ethics committee approvals have been obtained and details of the IRB/oversight body are included in the manuscript.
Yes
All necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived.
Yes
I understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).
Yes
I have followed all appropriate research reporting guidelines and uploaded the relevant EQUATOR Network research reporting checklist(s) and other pertinent material as supplementary files, if applicable.
Yes
Data Availability
Data used is present in the submitted article