ABSTRACT
Background It has been projected that there will be too few ventilators to meet demand during the COVID-19 (SARS CoV-2) pandemic. Ventilator sharing has been suggested as a crisis standard of care strategy to increase availability of mechanical ventilation. The safety and practicality of shared ventilation in patients is unknown. We designed and evaluated a system whereby one mechanical ventilator can be used to simultaneously ventilate two patients who have different lung compliances using a custom manufactured flow control valve to allow for individual adjustment of tidal volume and airway pressure for each patient.
Methods The system was first evaluated in a simulation lab using two human patient simulators under expected clinical conditions. It was then tested in an observational study of four patients with acute respiratory failure due to COVID-19. Two separately ventilated COVID-19 patients were connected to a single ventilator for one hour. This intervention was repeated in a second pair of patients. Ventilatory parameters (tidal volume, peak airway pressures, compliance) were recorded at five minute intervals during both studys. Arterial blood gases were taken at zero, thirty, and sixty minutes. The primary outcome was maintenance of stable acid-base status and oxygenation during shared ventilation.
Results Two male and two female patients, age range 32-56 yrs, participated. Ideal body weight and driving pressure were markedly different among patients. All patients demonstrated stable physiology and ventilation for the duration of shared ventilation. In one patient tidal volume was increased after 30 minutes to correct a respiratory acidosis.
Conclusions Differential ventilation using a single ventilator and a split breathing circuit with flow control valves is possible. A single ventilator could feasibly be used to safely ventilate two COVID-19 patients simultaneously as a bridge to full ventilatory support.
Summary Statement Not applicable.
Competing Interest Statement
The authors have declared no competing interest.
Clinical Trial
This study is an observational study to validate performance of a novel ventilator control valve. It is not a clinical trial.
Funding Statement
Internal funding
Author Declarations
All relevant ethical guidelines have been followed; any necessary IRB and/or ethics committee approvals have been obtained and details of the IRB/oversight body are included in the manuscript.
Yes
All necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived.
Yes
I understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).
Yes
I have followed all appropriate research reporting guidelines and uploaded the relevant EQUATOR Network research reporting checklist(s) and other pertinent material as supplementary files, if applicable.
Yes
Footnotes
Clinical trial number and registry UR: Not applicable.
Prior Presentations: Not applicable.
Funding Statement: Support was provided solely from institutional and/or departmental sources.
Conflicts of Interest: The authors declare no competing interests.
Data Availability
Data referred to in the manuscript will be made available if asked for.