ABSTRACT
Background Thermal damage to the esophagus is a risk from radiofrequency (RF) ablation of the left atrium for the treatment of atrial fibrillation (AF), with the most extreme type of thermal injury resulting in atrio-esophageal fistula (AEF), with a correspondingly high mortality rate. Various approaches have been developed to reduce esophageal injury, including power reduction, avoidance of greater contact-force, esophageal deviation, and esophageal cooling. One method of esophageal cooling involves direct instillation of cold water or saline into the esophagus during RF ablation. Although this method provides limited heat-extraction capacity, studies of it have suggested potential benefit.
Objective We sought to perform a meta-analysis of existing studies evaluating esophageal cooling via direct liquid instillation for the reduction of thermal injury.
Methods We reviewed Medline for existing studies involving esophageal cooling for protection of thermal injury during RF ablation. A meta-analysis was then performed using random effects model to calculate estimated effect size with 95% confidence intervals, with outcome of esophageal lesions, stratified by severity, as determined by post-procedure endoscopy.
Results A total of 9 studies were identified and reviewed. After excluding pre-clinical and mathematical model studies, 3 were included in the meta-analysis, totaling 494 patients. Esophageal cooling showed a tendency to shift lesion severity downward, such that total lesions did not show a statistically significant change (OR 0.6, 95% CI 0.15 to 2.38). For high grade lesions, a significant OR of 0.39 (95% CI 0.17 to 0.89) in favor of esophageal cooling was found, suggesting that esophageal cooling, even utilizing a low-capacity thermal extraction technique, reduces lesion severity from RF ablation.
Conclusions Esophageal cooling reduces lesion severity encountered during RF ablation, even when using relatively low heat extraction methods such as direct instillation of cold liquid. Further investigation of this approach is warranted.
Competing Interest Statement
EK declares equity interest in Attune Medical; MG, PS, CT, JG, and BC serve as Principal Investigators for studies of esophageal cooling, but receive no direct corporate compensation. SM has provided consulting services for Attune Medical. All other authors declare no relevant conflicts of interest.
Funding Statement
No external funding received.
Author Declarations
All relevant ethical guidelines have been followed and any necessary IRB and/or ethics committee approvals have been obtained.
Not Applicable
All necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived.
Not Applicable
Any clinical trials involved have been registered with an ICMJE-approved registry such as ClinicalTrials.gov and the trial ID is included in the manuscript.
Not Applicable
I have followed all appropriate research reporting guidelines and uploaded the relevant Equator, ICMJE or other checklist(s) as supplementary files, if applicable.
Yes
Footnotes
Conflict of interest: EK declares equity interest in Attune Medical; MG, PS, CT, JG, and BC serve as Principal Investigators for studies of esophageal cooling, but receive no direct corporate compensation. SM has provided consulting services for Attune Medical. All other authors declare no relevant conflicts of interest.
Funding: None
Data Availability
Data are available publicly and upon request.