Abstract
Background We evaluate the diagnostic performance of dried blood microsampling combined with a high-throughput microfluidic nano-immunoassay (NIA) for the identification of anti-SARS-CoV-2 Spike IgG seropositivity.
Methods We conducted a serological study among 192 individuals with documented prior SARS-CoV-2 infection and 44 SARS-CoV-2 negative individuals. Participants with prior SARS-CoV-2 infection had a long interval of 11 months since their qRT-PCR positive test. Serum was obtained after venipuncture and tested with an automated electrochemiluminescence anti-SARS-CoV-2 S total Ig reference assay, a commercial ELISA anti-S1 IgG assay, and the index test NIA. 109 participants from the positive cohort and 44 participants from the negative cohort also participated in capillary blood collection using three microsampling devices: Mitra, repurposed glucose test strips, and HemaXis. Samples were dried, shipped by regular mail, extracted, and measured with NIA.
Findings Using serum samples, we achieve a clinical sensitivity of 98·33% and specificity of 97·62% on NIA, affirming the high performance of NIA in participants 11 months post infection. Combining microsampling with NIA, we obtain a clinical sensitivity of 95·05% using Mitra, 61·11% using glucose test strips, 83·16% using HemaXis, and 91·49% for HemaXis after automated extraction, without any drop in specificity.
Interpretation High sensitivity and specificity was demonstrated when testing micro-volume capillary dried blood samples using NIA, which is expected to facilitate its use in large-scale studies using home-based sampling or samples collected in the field.
Funding Swiss National Science Foundation NRP 78 Covid-19 grant 198412 and Private Foundation of the Geneva University Hospital.
Evidence before this study Serological surveillance is of importance to better understand the evolution and spread of SARS-CoV-2 and adapt public health measures. We identified multiple studies conducting such serological surveys using decentralized collection of capillary blood, facilitating the logistics and reducing burden on participants and healthcare facilities. To perform the detection of anti-SARS-CoV-2 antibodies with a high-throughput and at low-cost, a microfluidic nano-immunoassay (NIA) was developed which requires ultra-low sample volumes and minimizes reagent consumption.
Added value of this study In this study we showed the possibility of combining capillary microsampling with NIA. We validated the use of NIA in serum samples obtained 11 months after infection and show the good clinical performance of the assay in samples with waning antibody titers. Using three different microsampling device, namely Mitra, repurposed glucose test strips, and HemaXis, we implemented a protocol using dried blood sample collection, shipping, extraction, and testing on the microfluidic assay. The sensitivity and specificity were measured and are presented when using the different microsampling devices.
Implications of all the available evidence We show that the performance of NIA is good when using serum samples, but also in combination with microsampling. Facilitated logistics and increased convenience of microsampling, together with high-throughput and low-cost testing on a microfluidic assay should facilitate the conduction of serological surveys.
Competing Interest Statement
The automated DBS extraction was performed using an instrument loaned free of charge by Gerstel AG. Gerstel AG did not take part in the decision to publish the study and did not have editorial control of the results presented.
Funding Statement
This work was supported by the SNF NRP 78 Covid-19 grant 198412 (SJM, IE, and BM), and a grant from the Private Foundation of the Geneva University Hospital (IE).
Author Declarations
I confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.
Yes
The details of the IRB/oversight body that provided approval or exemption for the research described are given below:
Ethical approval for this study was given by the Ethics Committee Geneva (Commission Cantonale d'Ethique de la Recherche sur l'etre humain (CCER) project numbers 2020-00516 and 2020-02323) and registered (NCT04329546) prior to initiation.
I confirm that all necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived, and that any patient/participant/sample identifiers included were not known to anyone (e.g., hospital staff, patients or participants themselves) outside the research group so cannot be used to identify individuals.
Yes
I understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).
Yes
I have followed all appropriate research reporting guidelines and uploaded the relevant EQUATOR Network research reporting checklist(s) and other pertinent material as supplementary files, if applicable.
Yes
Data Availability
All data included in the article and corresponding data will be made available as supplementary files with publication.