ABSTRACT
SARS-CoV-2 continues to develop new, increasingly infectious variants including delta and omicron. We evaluated the efficacy of the Abbott BinaxNOW Rapid Antigen Test against Reverse Transcription Polymerase Chain Reaction (“RT-PCR”) in 1054 pediatric participants presenting to a high-volume Coronavirus Disease 2019 (COVID-19) testing site while the delta variant was predominant. Participants were grouped by COVID-19 exposure and symptom status. RT-PCR demonstrated an overall prevalence of 5.2%. For all participants, sensitivity of the BinaxNOW was 92.7% (95% CI 82.4%-98.0%) and specificity was 98.0% (95% CI 97.0%-98.8%). For symptomatic participants, positive predictive value (PPV) was 72.7% (95% CI 54.5%-86.7%) and negative predictive value (NPV) was 99.2% (95% CI 98.2%-100%). Among asymptomatic participants, PPV was 71.4% (95% CI 53.7%-85.4%) and NPV was 99.7% (95% CI 99.0%-100%). Our reported sensitivity and NPV are higher than other pediatric studies, potentially because of higher viral load from the delta variant, but specificity and PPV are lower.
Importance The BinaxNOW rapid antigen COVID-19 test had a sensitivity of nearly 92% in both symptomatic and asymptomatic children when performed at a high-throughput setting during the more transmissible delta variant dominant period. The test may play an invaluable role in asymptomatic screening and keeping children safe in school.
Competing Interest Statement
Dr. Manabe reports institutional grants (to Johns Hopkins) from National Institutes of Health, Becton-Dickinson, Centers for Disease Control, and Baltimore City Health Department and institutional receipt (to Johns Hopkins) of equipment, materials, drugs, medical writing, gifts or other services from Hologic, Becton Dickinson, Roche, Cepheid, and ChemBio. Dr. Robinson reports institutional grants (toJohns Hopkins) from National Institutes of Health, Becton-Dickinson, Centers for Disease Control, and institutional receipt (to Johns Hopkins) of equipment, materials, drugs from Becton Dickinson. All other authors report no conflicts of interests.
Clinical Protocols
https://journals.asm.org/doi/10.1128/Spectrum.01008-21
Funding Statement
This study did not receive any funding. BinaxNOW test kits were provided by the Maryland Department of Health free of cost.
Author Declarations
I confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.
Yes
The details of the IRB/oversight body that provided approval or exemption for the research described are given below:
IRB of Johns Hopkins gave ethical approval for this work
I confirm that all necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived, and that any patient/participant/sample identifiers included were not known to anyone (e.g., hospital staff, patients or participants themselves) outside the research group so cannot be used to identify individuals.
Yes
I understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).
Yes
I have followed all appropriate research reporting guidelines and uploaded the relevant EQUATOR Network research reporting checklist(s) and other pertinent material as supplementary files, if applicable.
Yes
Footnotes
Abstract shortened and clarified
Data Availability
All data produced in the present work are available upon reasonable request to the authors