PT - JOURNAL ARTICLE AU - Higashimoto, Yuki AU - Ihira, Masaru AU - Kawamura, Yoshiki AU - Inaba, Masato AU - Shirato, Kazuya AU - Suzuki, Tadaki AU - Hasegawa, Hideki AU - Kageyama, Tsutomu AU - Doi, Yohei AU - Hata, Tadayoshi AU - Yoshikawa, Tetsushi TI - Dry loop–mediated isothermal amplification assay for detection of SARS-CoV-2 from clinical specimens AID - 10.1101/2020.09.29.20204297 DP - 2020 Jan 01 TA - medRxiv PG - 2020.09.29.20204297 4099 - http://medrxiv.org/content/early/2020/09/30/2020.09.29.20204297.short 4100 - http://medrxiv.org/content/early/2020/09/30/2020.09.29.20204297.full AB - Coronavirus disease 2019 (COVID-19) has had a major disease burden on many countries around the world. The spread of COVID-19 is anticipated to have a major impact on developing countries including African nations. To establish a point-of-care test for COVID-19, we developed a dry loop–mediated isothermal amplification (LAMP) method to detect severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) RNA. We carried out reverse transcription (RT)-LAMP using the Loopamp SARS-CoV-2 Detection kit (Eiken Chemical, Tokyo, Japan). The entire mixture except for the primers is dried and immobilized inside the tube lid. To determine the specificity of the kit, 22 viral genomes associated with respiratory infections, including the SARS coronavirus, were tested. No LAMP product was detected in reactions performed with RNA from these pathogens. The sensitivity of this assay, determined by either a real-time turbidity assay or colorimetric change of the reaction mixture, as evaluated by the naked eye or under illumination with ultraviolet light, was 10 copies/reaction. After the initial validation analysis, we analyzed 24 nasopharyngeal swab specimens collected from patients suspected to have COVID-19. Nineteen (79.2%) of the 24 samples were positive for SARS-CoV-2 RNA, as determined by real-time RT-PCR analysis. Using the Loopamp SARS-CoV-2 Detection kit, we detected SARS-CoV-2 RNA in 15 (62.5%) of the 24 samples. Thus, the sensitivity, specificity, positive predictive value, and negative predictive value of the Loopamp 2019-CoV-2 detection reagent kit were 94.0%, 96.0%, 95.9%, and 94.1%, respectively. The dry LAMP method for detection of SARS-CoV-2 RNA was fast and easy to use, solves the cold chain problem, and therefore represents a promising tool for diagnosis of COVID-19 in developing countries.Author summary Coronavirus disease 2019 (COVID-19) is a major public health problem around the world. A reliable point-of-care (POC) test for severe acute respiratory syndrome coronavirus 2 (SARS CoV-2) is urgently needed, especially in developing countries. The loop-mediated isothermal amplification (LAMP) method amplifies template nucleotides under isothermal conditions with high efficiency and specificity, both of which are major advantages for a POC test. In addition, because dry LAMP reagents can be stored at 4°C, it is suitable for use in developing countries.We evaluated the specificity and sensitivity of the Loopamp SARS-CoV-2 Detection kit (Eiken Chemical, Tokyo, Japan), a dry LAMP method for amplifying viral RNA. The initial validation study revealed that the method was highly specific and sensitive (lower detection limit: 10 copies/reaction). We then analyzed 24 nasopharyngeal swab specimens from patients suspected to have COVID-19. Using the Loopamp SARS-CoV-2 Detection kit, SARS-CoV-2 RNA was detected in 15 (62.5%) of the 24 samples. Compared with the standard real-time reverse transcription PCR, the sensitivity, specificity, positive predictive value, and negative predictive value of the Loopamp SARS-CoV-2 Detection kit were 78.9%, 100%, 100%, and 55.6%, respectively.Competing Interest StatementThe authors have declared no competing interest.Clinical TrialNo. HM19-493Funding StatementThe author(s) received no specific funding for this work.Author DeclarationsI confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.YesThe details of the IRB/oversight body that provided approval or exemption for the research described are given below:the institutional review board of Fujita Health UniversityAll necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived.YesI understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).YesI have followed all appropriate research reporting guidelines and uploaded the relevant EQUATOR Network research reporting checklist(s) and other pertinent material as supplementary files, if applicable.YesAll data generated or analysed during this study are included in this published article.