RT Journal Article SR Electronic T1 Agreement between arterial and capillary pH, pCO2 and lactate in patients in the emergency department JF medRxiv FD Cold Spring Harbor Laboratory Press SP 2020.06.11.20128223 DO 10.1101/2020.06.11.20128223 A1 Collot, V. A1 Malinverni, S. A1 Schweitzer, E. A1 Haltout, J. A1 Mols, P. A1 Bartiaux, M. YR 2020 UL http://medrxiv.org/content/early/2020/06/12/2020.06.11.20128223.abstract AB Study objective The primary objective of the study was a quantitative analysis to assess the mean difference and 95% confidence interval of the difference between capillary and arterial blood gas analyses for pH, pCO2 and lactate. Secondary objective was to measure the sensitivity and specificity of capillary samples to detect altered pH, hypercarbia and lactic acidosis.Methods Adults admitted to the ED for whom the treating physician deemed necessary an arterial blood gas analysis (BGA) were screened for inclusion. Simultaneous arterial and capillary samples were drawn for BGA. Agreement between the two methods for pH, pCO2 and lactate were studied with Bland-Altman bias plot analysis. Sensitivity, specificity, positive and negative predictive value as well as AUC were calculated for the ability of capillary samples to detect pH values outside normal ranges, hypercarbia and hyperlactatemia.Results 197 paired analyses were included in the study. Mean difference for pH, between arterial and capillary BGA was 0.0095, 95% limits of agreement were -0.048 to 0.067. For pCO2, mean difference was -0.3 mmHg, 95% limits of agreement were -8.5 to 7.9 mmHg. Lactate mean difference was -0.93 mmol/L, 95% limits of agreement were -2.7 to 0.8 mmol/L. At a threshold of 7.34 for capillary pH had 98% sensitivity and 97% specificity to detect acidemia; at 45.9 mmHg capillary pCO2 had 89% sensitivity and 96% specificity to detect hypercarbia. Finally at a threshold of 3.5 mmol/L capillary lactate had 66% sensitivity to detect lactic acidosis.Conclusion Capillary measures of pH, pCO2 and lactate can’t replace arterial measurements although there is high concordance between the two methods for pH and pCO2 and moderate concordance for lactate. Capillary blood gas analysis had good accuracy when used as a screening tool to detect altered pH and hypercarbia but insufficient sensitivity and specificity when screening for lactic acidosis.Competing Interest StatementThe authors have declared no competing interest.Funding StatementStefano Malinverni received an unconditional grant from Association pour l aide a la recherche medicale Andre VesaleAuthor DeclarationsI confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.YesThe details of the IRB/oversight body that provided approval or exemption for the research described are given below:Comite d'Ethique CHU Saint PierreAll necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived.YesI understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).YesI have followed all appropriate research reporting guidelines and uploaded the relevant EQUATOR Network research reporting checklist(s) and other pertinent material as supplementary files, if applicable.YesData supporting the manuscript will be made available upon request. https://doi.org/10.6084/m9.figshare.12464888.v1