PT - JOURNAL ARTICLE AU - Gultekin, David H. AU - Vaughan, J. Thomas AU - Shrivastava, Devashish TI - An experimental evaluation of the relationship between the induced radiofrequency heating near an implanted conductive medical device during MRI, scanner reported B1+rms, and scanner reported average transmit power AID - 10.1101/2024.03.04.24303732 DP - 2024 Jan 01 TA - medRxiv PG - 2024.03.04.24303732 4099 - http://medrxiv.org/content/early/2024/03/06/2024.03.04.24303732.short 4100 - http://medrxiv.org/content/early/2024/03/06/2024.03.04.24303732.full AB - Background Time-varying radiofrequency (RF) fields necessary to perform magnetic resonance imaging (MRI) may induce excessive heating near implanted conductive medical devices during MRI. Both time and space-averaged root mean square of the effective magnetic field (B1+rms) and whole-body average specific absorption rate (SAR) (average RF power per unit body weight) have been proposed as metrics to control the induced heating and avoid unintended thermal injury.Purpose To evaluate the relationship between the induced RF heating near an implanted conductive medical device, scanner-reported B1+rms, and scanner-reported RF power.Methods RF heating was measured near the electrodes of deep brain stimulation (DBS) lead placed in a gel phantom using fluoroptic temperature probes in a commercial 3T scanner during MRI. Four transmit and receive RF coil combinations were used, a circularly polarized head transmit and receive coil, a 20-channel head and neck, a 32-channel head, or a 64-channel head and neck receive-only coil with a whole-body transmit coil. RF heating was induced by running a 2D GRE sequence with two RF pulse types (fast and normal) with varying flip angles of 30°, 60°, and 90° and by turning the receive-only coils off and on. The scanner-reported B1+rms and RF power were recorded.Results Measurements show that the induced temperature change correlated linearly with both the scanner-reported B1+rms and RF power for each coil combination. However, the variation in the induced heating for various RF coil combinations appeared to be much larger for the scanner-reported B1+rms compared to the scanner-reported RF power.Conclusion Additional studies across other MR scanners are needed to better understand the full extent of the variation in the induced heating near implanted conductive devices as a function of the scanner-reported B1+rms and RF power to develop conservative and reliable patient labeling.Competing Interest StatementThe authors have declared no competing interest.Funding StatementThis study did not receive any fundingAuthor DeclarationsI confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.YesI confirm that all necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived, and that any patient/participant/sample identifiers included were not known to anyone (e.g., hospital staff, patients or participants themselves) outside the research group so cannot be used to identify individuals.YesI understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).YesI have followed all appropriate research reporting guidelines, such as any relevant EQUATOR Network research reporting checklist(s) and other pertinent material, if applicable.YesAll data produced in the present work are contained in the manuscript