RT Journal Article SR Electronic T1 Accurate 3D Positron Range Correction Method for Heterogeneous Material Densities in PET JF medRxiv FD Cold Spring Harbor Laboratory Press SP 2022.08.16.22278715 DO 10.1101/2022.08.16.22278715 A1 Li, Chong A1 Scheins, Jürgen A1 Tellmann, Lutz A1 Issa, Ahlam A1 Wei, Long A1 Shah, N. Jon A1 Lerche, Christoph YR 2022 UL http://medrxiv.org/content/early/2022/08/19/2022.08.16.22278715.abstract AB Objective The positron range is a fundamental, detector-independent physical limitation to special resolution in positron emission tomography (PET) as it causes a significant blurring of the reconstructed PET images. A major challenge for positron range correction methods is to provide accurate range kernels that inherently incorporate the generally inhomogeneous stopping power, especially at tissue boundaries. In this work, we propose a novel approach to generate accurate three-dimensional (3-D) blurring kernels both in homogenous and heterogeneous media to improve PET spatial resolution.Approach In the proposed approach, positron energy deposition was approximately tracked along straight paths, depending on the positron stopping power of the underlying material. The positron stopping power was derived from the attenuation coefficient of 511keV gamma photons according to the available PET attenuation maps. Thus, the history of energy deposition is taken into account within the range of kernels. Special emphasis was placed on facilitating the very fast computation of the positron annihilation probability in each voxel.Results Positron path distributions of 18F in low-density polyurethane were in high agreement with Geant4 simulation at an annihilation probability larger than 10−2∼10−3 of the maximum annihilation probability. The Geant4 simulation was further validated with measured 18F depth profiles in these polyurethane phantoms. The tissue boundary of water with cortical bone and lung was correctly modeled. Residual artifacts from the numerical computations were in the range of 1%. The calculated annihilation probability in voxels shows an overall difference of less than 20% compared to the Geant4 simulation.Significance The proposed method significantly improves spatial resolution for non-standard isotopes by providing accurate range kernels, even in the case of significant tissue inhomogeneities.Competing Interest StatementThe authors have declared no competing interest.Funding StatementThe research was supported by the Helmholtz-OCPC (Office of China Postdoc Council) Postdoc-Program through the Forschungszentrum Juelich GmbH. This work was also supported by the China National Natural Science Foundation (Grant Nos. U1932162).Author DeclarationsI confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.YesI confirm that all necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived, and that any patient/participant/sample identifiers included were not known to anyone (e.g., hospital staff, patients or participants themselves) outside the research group so cannot be used to identify individuals.YesI understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).YesI have followed all appropriate research reporting guidelines and uploaded the relevant EQUATOR Network research reporting checklist(s) and other pertinent material as supplementary files, if applicable.YesAll data produced in the present study are available upon reasonable request to the authors