RT Journal Article SR Electronic T1 Portable and label-free quantitative Loop-mediated Isothermal Amplification (qLAMP) for reliable COVID-19 diagnostics in 3 minutes: An Arduino-based detection system assisted by a pH microelectrode JF medRxiv FD Cold Spring Harbor Laboratory Press SP 2021.05.23.21256350 DO 10.1101/2021.05.23.21256350 A1 Moisés Alvarez, Mario A1 Bravo-González, Sergio A1 González-González, Everardo A1 Santiago, Grissel Trujillo-de YR 2021 UL http://medrxiv.org/content/early/2021/05/28/2021.05.23.21256350.abstract AB Loop-mediated isothermal amplification (LAMP) has been recently studied as an alternative method for cost-effective diagnostics in the context of the current COVID-19 pandemic. Recent reports document that LAMP-based diagnostic methods have a comparable sensitivity and specificity to that of RT-qPCR. We report the use of a portable Arduino-based LAMP-based amplification system assisted by pH microelectrodes for the accurate and reliable diagnosis of SARS-CoV-2 during the first 3 minutes of the amplification reaction. We show that this simple system enables a straightforward discrimination between samples containing or not containing artificial SARS-CoV-2 genetic material in the range of 10 to 10,000 copies per 50 µL of reaction mix. We also spiked saliva samples with SARS-CoV-2 synthetic material and corroborated that the LAMP reaction can be successfully monitored in real time using microelectrodes in saliva samples as well. These results may have profound implications for the design of real-time and portable quantitative systems for the reliable detection of viral pathogens including SARS-CoV-2.Competing Interest StatementWe are currently in the process of signing a collaboration agreement with Aim Brands, Llc.Funding StatementWe acknowledge funding from Tecnologico de Monterrey (Novus; Office of Technology Transfer - COVID-19 Emergency Funding). We are grateful and acknowledge support from Fundacion-FEMSA and form an altruist Tecnologico de Monterrey alumni who prefers to remain anonymous.Author DeclarationsI confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.YesThe details of the IRB/oversight body that provided approval or exemption for the research described are given below:The experimental protocol was approved on May 20th, 2020 by a named institutional committee (Alfa Medical Center, Research Committee; resolution AMCCI-TECCOVID-001).All necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived.YesI understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).YesI have followed all appropriate research reporting guidelines and uploaded the relevant EQUATOR Network research reporting checklist(s) and other pertinent material as supplementary files, if applicable.YesAll data relevant to the writing of this article is available upon request.