RT Journal Article SR Electronic T1 Direct detection of microorganisms responsible for the sexually transmitted infection on the mobile real-time PCR device without treating in advance JF medRxiv FD Cold Spring Harbor Laboratory Press SP 2021.03.02.21252726 DO 10.1101/2021.03.02.21252726 A1 Muraoka, Masaaki A1 Sohma, Kazunori A1 Kawaguchi, Osamu A1 Mizukoshi, Mikio YR 2021 UL http://medrxiv.org/content/early/2021/03/06/2021.03.02.21252726.abstract AB As WHO reported, four curable STIs-chlamydia, gonorrhoea, syphilis and trichomoniasis occur more than 1 million per each day globally almond 2016. For this reason, it is important to control these STIs, one of which is “to detect”. The general methods in order to detect STIs are nucleic acid amplification tests (NAATs). One of the reasons why NAATs are utilized in many tests is that it is possibly to be more sensitive than other test. However, there needs to treat extraction of nucleic acids in advance and amplify specific regions by NAATs, and hence it must take much labour and much time. In this work, for Chlamydia trachomatis (CT), Neisseria gonorrhoeae (NG) and Treponema pallidum (TP) which is each etiological agent of chlamydia, gonorrhoea and syphilis, we evaluate and propose “quicker and simpler” NAATs. Specifically, utilizing mobile real-time PCR device “PCR1100” and PCR reagent kit “KAPA3G Plant PCR Kit”, it was considered whether real-time direct PCR could be performed or not without treating DNA extraction in advance so-called “direct”.As a result, firstly, we established that real-time direct PCR could be performed in all of CT, NG, and TP, and moreover, each Ct value correlated with the concentration of each organism similarly to detection of genome DNA (each correlation coefficient R2 > 0.95). Moreover, each assay demonstrated a limit of detection (LOD) of the follows; CT was 10^0.86 = 7.24 IFU/reaction, NG was 10^-0.19 = 0.65 CFU/reaction, and TP was 10^1.4 = 25.1 organisms/reaction. However, it appeared the sensitivity was a little low, especially for CT and TP.Secondly, we found that even as without treating sample in advance, the time of detection was required more less 15 minutes at any of case, which was very quick compared with other current methods for real-time PCR. Additionally, compared with other commercial devices, it was easier to operate the PCR1100 device, for example, start, analysis of Ct value.In conclusion, the present study has demonstrated that it is possible for real-time direct PCR to perform with combination of the PCR1100 device and the PCR reagent kit in 3 kinds of microorganisms-CT, NG and TP. Furthermore, we propose “quicker and simpler” methods for NAATs, which it would not take labour and time. Further studies are needed in order to contribute to control STIs.Competing Interest StatementThe authors have declared no competing interest.Clinical TrialThis study do not contain any studies involving human participants and animals performed by any of the authors, so that is not involved in clinical trials.Funding StatementNo external funding was received to me this time.Author DeclarationsI confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.YesThe details of the IRB/oversight body that provided approval or exemption for the research described are given below:I did not have to be obtained necessary IRB and/or ethics committee approvals in this time.All necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived.YesI understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).YesI have followed all appropriate research reporting guidelines and uploaded the relevant EQUATOR Network research reporting checklist(s) and other pertinent material as supplementary files, if applicable.YesThe data that support the findings of this study are available from the corresponding author, upon reasonable request.