PT - JOURNAL ARTICLE AU - Ruscio, Maurizio AU - D’Agnolo, Elisa AU - Belgrano, Anna AU - Plebani, Mario AU - Lippi, Giuseppe TI - Analytical assessment of Beckman Coulter Access anti-SARS-CoV-2 IgG immunoassay AID - 10.1101/2020.11.05.20226555 DP - 2020 Jan 01 TA - medRxiv PG - 2020.11.05.20226555 4099 - http://medrxiv.org/content/early/2020/11/07/2020.11.05.20226555.short 4100 - http://medrxiv.org/content/early/2020/11/07/2020.11.05.20226555.full AB - Background The approach to diagnosing, treating and monitoring severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) infection relies strongly on laboratory resources, with serological testing representing the mainstay for studying the onset, nature and persistence of humoral immune response. This study was aimed at evaluating the analytical performance of the novel Beckman Coulter anti-SARS-CoV-2 IgG chemiluminescent immunoassay.Methods This analytical assessment encompassed the calculation of intra-assay, inter-assay and total imprecision, linearity, limit of blank (LOB), limit of detection (LOD), functional sensitivity, and comparison of anti-SARS-CoV-2 antibodies values obtained on paired serum samples using DiaSorin Liaison SARS-CoV-2 S1/S2 IgG and Roche Elecsys Anti-SARS-CoV-2 total antibodies. Diagnostic performance was also tested against results of molecular testing on nasopharyngeal swabs, collected over the previous 4 months.Results Intra-assay, inter-assay and total imprecision of Beckman Coulter anti-SARS-CoV-2 IgG were between 4.3-4.8%, 2.3-3.9% and 4.9-6.2%, respectively. The linearity of the assay was excellent between 0.11-18.8 antibody titers. The LOB, LOD and functional sensitivity were 0.02, 0.02 and 0.05, respectively. The diagnostic accuracy (area under the curve; AUC) of Beckman Coulter anti-SARS-CoV-2 IgG compared to molecular testing was 0.87 (95% CI, 0.84-0.91; p<0.001) using manufacturer’s cut-off, and increased to 0.90 (95% CI, 0.86-0.94; p<0.001) with antibody titers. The AUC was non-significantly different from that of Roche Elecsys Anti-SARS-CoV-2, but was always higher than that of DiaSorin Liaison SARS-CoV-2 S1/S2 IgG. The correlation of Beckman Coulter Access SARS-CoV-2 IgG was 0.80 (95% CI, 0.75-0.84; p<0.001) with Roche Elecsys Anti-SARS-CoV-2 and 0.72 (95% CI, 0.66-0.77; p<0.001) with DiaSorin Liaison SARS-CoV-2 S1/S2 IgG, respectively.Conclusions The results of this analytical evaluation of Beckman Coulter Access anti-SARS-CoV-2 IgG suggests that this fully-automated chemiluminescent immunoassay represents a valuable resource for large and accurate seroprevalence surveys.Competing Interest StatementThe authors have declared no competing interest.Funding StatementNo fundingAuthor DeclarationsI confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.YesThe details of the IRB/oversight body that provided approval or exemption for the research described are given below:Hospital of TriesteAll necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived.YesI understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).Yes I have followed all appropriate research reporting guidelines and uploaded the relevant EQUATOR Network research reporting checklist(s) and other pertinent material as supplementary files, if applicable.YesData are availabile from the authors upon request.