Abstract
Aims Severity assessment for coarctation of the aorta (CoA) is challenging due to concomitant morphological anomalies (complex CoA) and inaccurate Doppler-based indices. Promising diagnostic performance has been reported for the continuous flow pressure gradient (CFPG), but it has not been studied in complex CoA. Our objective was to characterize the effect of complex CoA and associated hemodynamics on CFPG in a clinical cohort.
Methods and Results Retrospective analysis identified discrete juxtaductal (n=25) and complex CoA (n=43; transverse arch and/or isthmus hypoplasia) patients with arm-leg systolic blood pressure gradients (BPG) within 24 hours of echocardiography for comparison to BPG by conventional Doppler indices (simplified Bernoulli equation and modified forms correcting for proximal kinetic energy and/or recovered pressure). Results were interpreted using the current CoA guideline (BPG ≥20 mmHg) to compare diagnostic performance indicators including receiver operating characteristic curves, sensitivity, specificity, and diagnostic accuracy, among others. Echocardiography Z-scored aortic diameters were applied with computational stimulations from a preclinical CoA model to understand aspects of the CFPG driving performance differences.
Diagnostic performance was substantially reduced from discrete to complex CoA for conventional Doppler indices calculated from patient data, and by hypoplasia and/or long segment stenosis in simulations. In contrast, diagnostic indicators for the CFPG only modestly dropped for complex vs discrete CoA. Simulations revealed differences in performance due to inclusion of the Doppler velocity index and diastolic pressure half-time in the CFPG calculation.
Conclusion CFPG is less affected by aortic arch anomalies co-existing with CoA when compared to conventional Doppler indices.
Introduction
Coarctation of the aorta (CoA) is a constriction of the thoracic aorta and is one of the most common congenital cardiovascular defects1–3. While surgery is the optimal treatment in most infants, 5-30% of treated patients develop re-coarctation requiring reintervention4 and > 90% develop hypertension after 50 years of age5. Guidelines for intervention include a transcatheter peak-to-peak CoA gradient (BPGpp) ≥20 mmHg that is often regarded as hemodynamically significant6.
Despite allowing for direct measurement of the gradient, catheterization is an invasive procedure with potentially unwanted risks7. BPGpp can also be estimated via cuffs placed on the arms and legs. However, this is difficult in patients clinically due to children who may be upset, crying or scared during measurement, and/or inappropriately sized cuffs8–10. Moreover, CoA patients frequently have anomalies in their arterial anatomy that limit accurate determination of BPGpp via BP cuffs. For example, the left subclavian artery is hypoplastic in many patients, and branches distal to the coarctation in ∼5% of patients11. Non-invasive alternatives using continuous wave Doppler are therefore commonly applied in echocardiography to estimate the BPGpp12. This often involves the systolic gradient estimated from peak Doppler velocity using the simplified Bernoulli equation (SBE), or modified forms correcting for proximal kinetic energy (i.e., modified Bernoulli equation: MBE) and/or recovered pressure (SBE-RP)13. These conventional Doppler estimates are known to have poor agreement with catheterization14. Sources of disagreement include local discrepancy in aortic dynamic distensibility, simplification of the Bernoulli equation, and morphological variability, which can result in a substantial Doppler misclassification1,15. This misclassification is a critical barrier to assessing treatment modifications and translation since interpretation of new thresholds is best done using tools with a high degree of accuracy and diagnostic performance.
In contrast to tools exclusively focused on the systolic pressure gradient, we recently reported on a novel Doppler-based tool that accounts for the diastolic pressure gradient, which we termed the continuous flow pressure gradient (CFPG)15. Importantly, the CFPG was recently reported as an independent predictor of hemodynamic severity in the setting of discrete CoA devoid of complications found in complex CoA1. The physiologic underpinnings of a pronounced diastolic gradient can be explained by the prolonged recoil (i.e. capacitance) of the proximal aorta in response to the flow-limiting stenosis that leads to antegrade diastolic run-off with a typical sawtooth velocity decay1. As mentioned above, CoA in patients often includes hypoplasia in one or more arch segments presenting in concert with discrete CoA, or a long segment stenosis resulting in additional sites of local blood flow acceleration (i.e., complex CoA). The objective of the current study was to assess the diagnostic performance of the CFPG relative to conventional Doppler indices in a pediatric cohort of discrete and complex CoA cases. We further set out to determine the aspects of the CFPG that are driving differences in performance relative to conventional Doppler-based indices, which was conducted using Z-scored aortic diameters from the pediatric cohort in concert with representative computational hemodynamic simulations based on a preclinical model of CoA.
Methods
Study design and patient involvement
To identify the most relevant research topics and meaningful outcomes, the study was specially designed with the help of colleagues from the Division of Pediatric Cardiothoracic Surgery at Children’s Wisconsin and Herma Heart Institute, but no direct patient involvement was considered due to retrospective nature of the study design. After exempt determination by the IRB Board of Children’s Wisconsin, analysis of 259 CoA patients of either sex identified 25 cases with discrete juxtaductal CoA and 43 with complex CoA (transverse arch and/or isthmus hypoplasia), as well as contemporaneous arm-leg systolic blood pressure gradients via sphygmomanometry within 24 hours of echocardiography. Despite potential limitations8–10, sphygmomanometry was interpreted as a ground-truth for Doppler-based estimates of the gradient from CoA since catheter-based BP gradients were not available. Patients with patent ductus arteriosus, more than mild aortic valve insufficiency, other hemodynamically significant congenital heart diseases, low cardiac output state, and abnormal ventricular systolic function on echocardiography examination were excluded. Table 1 summarizes patient characteristics.
Echocardiographic measurements included conventional peak instantaneous pressure gradient estimation by SBE, MBE and SBE-RP16. An estimate of the pressure gradient during diastole was also calculated using diastolic flow continuation as characterized by the CFPG discussed in detail elsewhere1. All measurements were quantified from 2D and spectral Doppler images with early diastolic velocity identified at the end of the T wave1. Details related to echo quantification methods are summarized in Supplemental Table 1.
Computational assessment of Doppler-derived index performance
Computational models were constructed using a pre-clinical rabbit model mimicking CoA in patients17,18 since measurements necessary to replicate physiology and aortic morphology beyond the coarctation region were not available in the echocardiography data from pediatric patients retrospectively studied. This approach allowed for understanding of hemodynamic factors impacting and/or potentially limiting the utility of the Doppler-based indices studied, including CFPG. Arterial segments were generated in SimVascular (simtk.org) using a control rabbit MRI dataset obtained following Institutional Animal Care and Use Committee approval2. The aortic diameter was then quantified at various locations throughout the thoracic aorta. This step was important to identify the normal population sized as baseline and realistically replicate discrete and complex CoA characteristics by matching Z-scores from computational models to the clinical datasets.
Model construction involved centerline detection, segmentation using a semi-automated level set method19, and lofting to create the blood flow domain. An in-house code was then developed to generate local modifications with the guidance of a pediatric cardiovascular surgeon by smoothly scaling aortic cross-sectional area according to shape functions20 representing different types of complex CoA including long-segment CoA, arch and isthmus hypoplasia (Supplemental Figure 1). The resulting morphological properties replicated Z-scores observed at the equivalent locations for each of the pediatric CoA patients21. A total of 68 computational aortas (25 with discrete CoA and 43 with complex CoA) were created. The aorta models were then integrated with arch branches (Supplemental Figure 2) for hemodynamic simulations with one inlet at the Sino-tubular junction and outlets at each of the head and neck arteries.
One-dimensional computational reduced order models (ROM) were generated in SimVascular (simvascular.github.io/) from the 3D CoA models22. The ROM approach was chosen for fast determination of hemodynamic results used to calculate Doppler indices (Supplemental Table 1) and automating the simulation process as compared to a semi-automated computationally intensive process required for 3D simulations with additional spatiotemporal information not necessary for the purpose of calculating Doppler-derived indies. A 3-element Windkessel model was implemented at each outlet to realistically replicate downstream pressure-flow relationships23 based on empirical measurements of pressure and flow from rabbits with discrete CoA. Windkessel model parameters were then adjusted to represent hemodynamics for complex CoA cases (see details in Supplemental Table 2). Data from discrete CoA rabbits with 42-96% area obstruction were used to replicate clinical BPGpp18 and the influence of CoA on thickness and stiffness after remolding of the wall that were included in computational simulations (see supplementary materials).
Statistical analysis
Hypoplasia proximal to the CoA was quantitatively characterized using Z-score criteria24. Arch and isthmus hypoplasia was considered significant for Z-scores <-3.0 (i.e., dimensions in the lowest 0.1% of the population). Pearson’s correlation and linear regression were used to characterize the relationship between Doppler-based estimates and simulated pressure gradients. Adjusted R-squared values quantified the degree of scatter for each CoA severity and morphology studied. Diagnostic performances were compared via receiver operative characteristic (ROC) curves and relative to the current intervention guideline17,25 (i.e., BPGpp ≥20 mmHg). The optimum cutoffs for Doppler indices were identified by maximizing the sum of sensitivity and specificity, (i.e.; the Youden index26) and the area under the ROC curves (AUC) was compared using two-tailed student’s t-test with 5 and 10% significance levels. Finally, sensitivity, specificity, positive predictive values (PPV), negative predictive value (NPV), diagnostic odds ratio, and diagnostic accuracy were also quantified as measures of diagnostic performance. Reliability of the echo-based measurements used in calculating metrics including, SBE, MBE, SBE-RP, and indices used in the calculation of the CFPG including the diastolic pressure half-time (dPHT) and Doppler velocity index (DVI) was investigated using intraclass correlation coefficient (ICC) through SPSS software. Two observers were trained to quantify echo-based indices on a random subgroup of simulation and pediatric CoA patient datasets (n=10). Each metric was quantified in triplicate for inter-observer variability with median values used for intra-observer variability analysis27.
Results
Doppler quantification from the cohort of pediatric CoA patients showed decreases in diagnostic performance moving from discrete to complex CoA morphologies, where correlation to proximal acceleration was evident. For example, the diagnostic accuracy was 0.88, 0.87, 0.79 for SBE, MBE, and SBE-RP indices from discrete CoA cases. These values dropped to 0.69, 0.59, and 0.59 for complex CoA cases, respectively. The diagnostic accuracy of CFPG was 0.96 for discrete CoA morphologies, and modestly dropped to 0.88 for complex CoA data sets. A detailed list of diagnostic performance indicators is provided in Table 2.
Similar patterns of reduced diagnostic performance were observed from the ROC analysis and AUCs. Specifically, the AUC was 0.91, 0.91, and 0.90 for conventional Doppler indices (i.e., SBE, MBE, and SBE-RP), which all significantly dropped to 0.80, 0.83, and 0.85, respectively (p-values <0.01, Table 2). On the other hand, the AUC for CFPG remained high at 0.98-0.99 despite additionally narrowing proximal to the coarctation for complex CoA cases (Table 2).
Reliability of the echo-based measurements used in calculating CFPG as quantified by ICC index revealed ≤9.8 and 11.2% differences among inter and intra-observer measurements, respectively. ICC index also confirmed good to excellent inter and intra-observer agreement in the conventional Doppler indices quantified. The 95% confidence interval for ICC was [0.86,0.98] for single subject variability and [0.91,0.99] on average, which is referred as excellent reliability according to Cicchetti et al28 and good according to Koo et al27.
Before assessing the diagnostic performance of each Doppler-based index quantified using echo data from the retrospective cohort of pediatric CoA patients via simulation, we first confirmed morphological similarity via Z-scores. Z-scores at the proximal transverse arch, distal transverse arch and aortic isthmus region were −3.22 ±0.81, −3.01 ±0.72, and −3.23 ±1.50 in the cohort of pediatric CoA patients, and corresponding values from computational models were −3.18 ±1.18, 3.11 ±0.52, and 3.17 ±1.68, which were not significantly different (p-values >0.1).
When compared to a discrete CoA, predictive values from conventional Doppler indices from ROM simulations were substantially reduced by the addition of hypoplasia and/or long segment stenosis in the arch. Specifically, the diagnostic accuracy was 0.63, 0.82, 0.80 for SBE, MBE, and SBE-RP indices for simulations with discrete CoA configurations. These values dropped to 0.48, 0.80, and 0.73 for simulations of complex CoA models, respectively. In contrast, the diagnostic accuracy of CFPG was 0.90 for simulations of discrete CoA morphologies, and only modestly dropped to 0.83 for simulations of complex CoA. A detailed list of predictive values is provided in Table 3.
A similar pattern of reduced diagnostic performance was observed in AUCs calculated from ROC curves of Doppler indices determined by discrete CoA simulations (Figure 1, Table 3). Specifically, the AUC was 0.90, 0.90, and 0.89 for conventional Doppler indices (i.e., SBE, MBE, and SBE-RP) calculated from simulations, which significantly dropped to 0.83, 0.86, and 0.83, respectively (p-values <0.01, Table 2). Moving from discrete to complex CoAs, ROC analysis showed a pronounced shift towards non-diagnostic line (identity line, bottom left-to upper-right) for simulations of conventional Doppler indices. On the other hand, ROC curves obtained by CFPG calculated from simulations were relatively unchanged.
For simulations of discrete CoA, regression analysis resulted in R2 = 0.88 for CFPG and 0.74-0.76 for conventional Doppler indices (Figure 2). Simulations of complex CoAs revealed the impact concomitant anomalies had on resulting flow patterns. For example, hemodynamic alterations including proximal acceleration and increased jet velocity were observed resulting in a higher degree of scatter (Figure 2). These factors decreased R2 values for conventional Doppler indices from simulations to 0.51-0.62, but to a lesser extent for CFPG calculated from simulations (0.78), indicating that CFPG outperformed conventional Doppler indices for simulations of both discrete and complex CoA.
Discussion
Suboptimal agreement with gold-standard catheterization represents a major limitation of conventional Doppler-derived indices used for non-invasive CoA severity assessment14,29. Although diastolic continuation of flow and estimation of its associated pressure gradient using the CFPG index previously showed strong correlation with measured values in a group of 25 discrete CoA patients1, its performance relative to conventional Doppler-derived indices had not been assessed in complex CoA morphologies. The main finding of the current study is that the CFPG was more consistent than any of the other Doppler-derived indices studied in the setting of complex CoA in a cohort of pediatric patients. It is also important to note that our results indicate the CFPG better correlates with arm-leg pressure gradients than the other Doppler-derived indices studied.
Performance differences among our novel CFPG method and conventional Doppler-derived indices can be understood relative to the contribution of local stenoses beyond discrete CoA using the simulations recounted above where physiologic boundary conditions were used to computationally investigate the diagnostic performance of CFPG and three conventional Doppler indices (SBE, MBE, and SBE-RP). Special care was taken to replicate morphological anomalies seen in complex forms of pediatric CoA patients by implementing similar Z-scores in proximal transverse arch, distal transverse arch, and aortic isthmus regions characterizing arch hypoplasia relative to the normal population24 (Table 1).
Differences in performance made clear by these simulations are rooted in the fact that the CFPG includes several additional sub-indices described in Supplemental Table 1, namely the DVI and dPHT, which consider the hemodynamic changes resulting in complex CoA cases. For example, in complex CoA, the DVI (proximal to jet peak velocity ratio) can increase due to proximal acceleration from a hypoplastic arch or isthmus (Figure 3, left). This leads to underestimation of CoA (Figure 2, B and C), and leads to discrepancy between discrete and complex CoA cases. However, the dPHT, which represents the time to half of the early diastolic pressure (Vd to interval on the spectral doppler image, see Supplemental Table 1) accounts for the prolonged recoil of the proximal aorta and associated antegrade diastolic flow. This leads to consistent behavior of the dPHT in the setting of additional proximal stenoses in the form of a hypoplastic arch and/or isthmus (Figure 3, right).
Consequently, although a mixed behavior is possible due to the DVI and dPHT having opposite effects on the CFPG as indicated in Supplemental Figure 3, the CFPG is still relatively insensitive to proximal acceleration in the clinical ranges of complex CoA studied. Specifically, CFPG increases with dPHT making it more sensitive to additional proximal narrowing while, on the other hand, it decreases with DVI as is also seen in MBE and SBE-RP. Therefore, collectively the CFPG performance tends to be less affected by the proximal acceleration as suggested in the AUCs shown in Figure 1 and values reported in Table 3.
Pronounced overestimation of the peak instantaneous pressure gradient commonly reported in the literature for conventional Doppler-derived indices is further confirmed in the current work, leading to superficial sensitivity of 1.00 for SBE compromising specificity. Conversely, sensitivity values <0.14 will lead to a superficial diagnostic odds ratio (DOR) of infinity that positively correlates with balanced sensitivity and specificity under normal circumstance. This imbalance between sensitivity and specificity is particularly challenging in translational studies were intervention guidelines (e.g., CoA gradient) are ideally identified using invasive high-fidelity pressure measurements techniques such as catheterization, but results are being used in clinical setting based on non-invasive estimations such as echocardiography. Hence, the improved sensitivity-specificity balance observed in CFPG (Table 3) is promising evidence that this index can be of more potential value translationally. Particularly, in the setting complex CoA morphology, CFPG continues to deliver excellent balance between high sensitivity and specificity that is also evident in DORs when compared to conventional Doppler indices. In clinical practice, CFPG can be interpreted relative to the arm-leg cuff gradient identified using the current dataset. Table 4 summarized CFPG cutoffs for a range of clinically important CoA gradients.
The current results should be interpreted relative to several limitations. For example, literature suggests the arm-leg gradient may not be a strong predictor of invasive gradient in CoA patients1,30. Therefore, we expect the CFPG cutoffs obtained here may be different if sphygmomanometry data are compared to gold-standard catheterization data, which were not available for the cohort of pediatric CoA patients studied. Additionally, the cohort studied represented pediatric patients. Hence, it remains to be determined if the relationships presented in Tables 2 and 4 are also generalizable to older CoA patients. Complex morphologies were generated in computational representations of rabbit aortas by replicating Z-scores seen clinically at the thoracis aorta for CoA patients. This is in contrast to using measurements from the echo data within the CoA region of pediatric patients that would need to be coupled with assumptions related to blood flow distributions and dimensions of the ascending aorta and its branches. Considering these limitations, clinical validation of CFPG in preventing postsurgical complications including hypertension has yet to be studied in future work.
Data Availability
All data produced in the present study are available upon reasonable request to the authors.
Disclosure
The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Funding
NIH: R01-HL142955 and R15-HL096096
Supplementary materials
Supplementary materials are available online.
Acknowledgements
The authors gratefully acknowledge the technical support from Leanne Harman, Lindsey Kalvin, Jamasp Azarnoosh, and Chalani Ellepola.
Footnotes
Emails: ghorbannia{at}outlook.com, aspearman{at}childrenswi.org, shahd.sawalhi{at}marquette.edu, rwoods{at}mcw.edu, mehdi.maadooliat{at}marquette.edu, jladisa{at}mcw.edu
Abbreviations
- AA
- Ascending aorta
- BP
- Blood pressure
- CFPG
- Continuous flow pressure gradient
- CoA
- Coarctation of the aorta
- DBP
- Diastolic blood pressure
- DVI
- Doppler velocity index
- MBE
- Modified Bernoulli equation
- PC-MRI
- Phase contrast magnetic resonance image
- ROM
- Reduced order model
- SBE
- Simplified Bernoulli equation
- SBE-RP
- SBE modified for recovered pressure
- SBP
- Systolic blood pressure