Abstract
Objective To evaluate the accuracy, reliability, clinical utility, and usability of HeraBEAT, a wireless fetal and maternal heart rate monitor (HBM) when used by clinicians and pregnant women to monitor fetal heart rate (FHR).
Methods We recruited women aged 18 years or older with a singleton pregnancy of ≥12 weeks gestation. FHR recordings were performed using the HBM and cardiotocography (CTG) to determine comparative accuracy. The HBM was then used by clinicians and participants in the antenatal clinic with the latter then using the device unassisted to record at home. The women rated the HBM using the System Usability Scale (SUS).
Results A total of 81 participants provided 126 recordings for analysis. The accuracy of the HBM was excellent compared with CTG, with limits of agreement (95%) between −1.5 and +0.9 beats per minute (bpm) and a mean difference of −0.29 bpm. The FHR was detected on 100% of occasions by clinicians (52 recordings) and participants when used in the clinic (42 recordings) and at home (32 recordings). Home users took an average of 1.1 minutes to detect the FHR and recorded a continuous trace of >1 minute in 94% of occasions, with an average total trace time of 4.4 minutes. The FHR trace was deemed to be clinically useful in 100% of clinician recordings and 97% of home recordings. There was no effect from body mass index, gestational age, pregnancy history, or placental position. The HBM ranked in the 96–100th percentile on the SUS for usability and learnability.
Conclusions The HBM was accurate and easy for clinicians and participants to use. The data recorded at home was equivalent to that obtained in the clinic using current assessment protocols for low-risk pregnancies, allowing the device to be used in telehealth consultations.
Clinical Trial Registration Australian New Zealand Clinical Trial Registry, https://www.anzctr.org.au ACTRN12620000739910.
Competing Interest Statement
Authors DS and JCh are shareholders of HeraMED. PP declares financial interest of a beneficial association with an entity owning shares and options in HeraMED Pty Ltd. No other author declares a conflict of interest.
Clinical Trial
ACTRN12620000739910
Funding Statement
The study was supported by PHI Research Group (not for profit) who were responsible for the study design, data collection and preparation of the manuscript. Joondalup Health Campus provided infrastructure support and IT services in-kind to the PHI research group. The HeraBEAT devices used in this study were loaned by HeraMED Pty Ltd (HeraMED, Netanya, ISRAEL(. NB and JC received consulting fees for statistical analysis from PHI research group funds.
Author Declarations
I confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.
Yes
The details of the IRB/oversight body that provided approval or exemption for the research described are given below:
Ethical approval was obtained from the Ramsay Health Care Human Research Ethics Committee of Victoria and New South Wales (reference number 2020-005).
All necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived.
Yes
I understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).
Yes
I have followed all appropriate research reporting guidelines and uploaded the relevant EQUATOR Network research reporting checklist(s) and other pertinent material as supplementary files, if applicable.
Yes
Data Availability
Unidentified data is available from the corresponding author for analysis dependent upon obtaining reciprocal ethical approval for sharing.