Abstract
Background Performances of rapid antigen diagnostic tests (Ag-RDTs) with nasal self-sampling, and oropharyngeal plus nasal (OP-N) self-sampling, in the Omicron period are unknown.
Methods Prospective diagnostic accuracy study among 6,497 symptomatic individuals aged >16 years presenting for SARS-CoV-2 testing at three test-sites. Participants were sampled for RT-PCR (reference test) and received one Ag-RDT to perform unsupervised with either nasal self-sampling (during the emergence of Omicron, and after Omicron share was >90%, phase-1) or with OP-N self-sampling (in a subsequent phase-2; Omicron share >99%). The evaluated tests were Acon Flowflex (Flowflex; phase-1 only), MP Biomedicals (MPBio), and Siemens-Healthineers Clinitest (Clinitest).
Findings During phase-1, 45% of Flowflex, 29% of MPBio, and 35% of Clinitest participants were confirmatory testers (previously tested positive by a self-test at own initiative). Overall sensitivities with nasal self-sampling were 79.0% (95% CI: 74.7-82.8%) for Flowflex, 69.9% (65.1-74.4%) for MPBio, and 70.2% (65.6-74.5%) for Clinitest. Sensitivities were substantially higher in confirmatory testers (93.6%, 83.6%, and 85.7%, respectively) than in those who tested for other reasons (52.4%, 51.5%, and 49.5%, respectively). Sensitivities decreased by 6.1 (p=0.16 by Chi-square test), 7.0 (p=0.60), and 12.8 (p=0.025) percentage points, respectively, when transitioning from 29% to >95% Omicron. During phase-2, 53% of MPBio, and 44% of Clinitest participants were confirmatory testers. Overall sensitivities with OP-N self-sampling were 83.0% (78.8%-86.7%) for MPBio and 77.3% (72.9%-81.2%) for Clinitest. Comparing OP-N to nasal sampling, sensitivities were slightly higher in confirmatory testers (87.4% and 86.1%, respectively), and substantially higher in those testing for other reasons (69.3% and 59.9%, respectively).
Interpretatio Sensitivities of three Ag-RDTs with nasal self-sampling decreased during Omicron emergence but was only statistically significant for Clinitest. Sensitivities were substantially influenced by the proportion of confirmatory testers. Addition of oropharyngeal to nasal self-sampling improved sensitivities of MPBio and Clinitest.
Funding Dutch Ministry of Health, Welfare, and Sport.
Evidence before this study SARS-CoV-2 rapid antigen diagnostic tests (Ag-RDTs) require no or minimal equipment, provide a result within 15-30 minutes, and can be used in a range of settings including for self-testing at home. Self-testing may potentially lower the threshold to testing and allows individuals to obtain a test result quickly and at their own convenience, which could support the early detection of infectious cases and reduce community transmission. Real world evidence on the performance of unsupervised nasal and oropharyngeal plus nasal (OP-N) self-sampling in the Omicron variant period is needed to accurately inform end-users and policymakers. Therefore, we conducted a large prospective diagnostic accuracy study of three commercially available Ag-RDTs with self-sampling (the Acon Flowflex test, the MP Biomedicals test, and the Siemens-Healthineers Clinitest) during and after the emergence of Omicron using RT-PCR as the reference standard. Our aims were to evaluate whether the accuracies of Ag-RDTs with nasal self-sampling changed over time with the emergence of Omicron; and to determine whether addition of oropharyngeal to nasal self-sampling with the same swab yielded higher diagnostic accuracies.
What this study adds The large comprehensive study was conducted in almost 6,500 participants with symptoms when presenting for routine SARS-CoV-2 testing at three public health service COVID-19 test-sites in the Netherlands. During the study, conducted between 21 December 2021 and 10 February 2022, the percentage of the Omicron variant in samples from the national SARS-CoV-2 pathogen surveillance increased from 29% in the first week to 99% in the last week of the study. The period during which the Omicron variant was dominant was divided into a nasal sampling phase (phase-1; Omicron present in >90% of surveillance samples) and an OP-N sampling phase (phase-2; Omicron share was >99%). In phase-1, 45% of Flowflex, 29% of MPBio, and 35% of Clinitest participants visited the test-site because of a positive self-test (confirmatory testers). Overall sensitivities with nasal self-sampling were 79.0% (95% CI: 74.7-82.8%) for the Flowflex, 69.9% (65.1-74.4%) for the MPBio, and 70.2% (65.6-74.5%) for the Clinitest Ag-RDT. Sensitivities were 94%, 84%, and 86%, respectively, for confirmatory testers, and 52%, 52%, and 50%, respectively, for those who had other reasons for getting tested. Sensitivities were 87.0% (79.7-92.4%), 83.1% (72.9-90.7%), and 80.0% (51.9-95.7%), respectively, in the first week, and decreased by 6.1 (p=0.16 by Chi-square test), 7.0 (p=0.60), and 12.8 (p=0.025) percentage points in the final week of the study. In Phase-2, 53% of MPBio and 44% of Clinitest participants were confirmatory testers. Overall sensitivities with OP-N self-sampling were 83.0% (78.8%-86.7%) for MPBio and 77.3% (72.9%-81.2%) for Clinitest. When comparing OP-N to nasal sampling, sensitivities were slightly higher in confirmatory testers (87.4% and 86.1%, respectively), and substantially higher in those testing for other reasons (69.3% and 59.9%).
Implications of all the available evidence The sensitivities of three commercially available Ag-RDTs performed with nasal self-sampling decreased during the emergence of Omicron, but this trend was only statistically significant for Clinitest. Addition of oropharyngeal to nasal self-sampling improved the sensitivity of the MPBio and Clinitest, most notably in individuals who visited the test-site for other reasons than to confirm a positive self-test. Based on these findings, the manufacturers of MPBio and Clinitest may consider extending their instructions for use to include combined oropharyngeal and nasal sampling, and other manufacturers may consider evaluating this as well.
Competing Interest Statement
The authors have declared no competing interest.
Funding Statement
This study was funded by the Dutch Ministry of Health, Welfare, and Sport. The funder had no role in design; collection, analysis, and interpretation of data; writing and decision to submit the paper for publication.
Author Declarations
I confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.
Yes
The details of the IRB/oversight body that provided approval or exemption for the research described are given below:
Ethical approval Not required because the study was judged by the METC Utrecht to be outside the scope of the Dutch Medical Research Involving Human Subjects Act (protocol No 21-818 /C). All participants signed an informed consent form before any study procedure.
I confirm that all necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived, and that any patient/participant/sample identifiers included were not known to anyone (e.g., hospital staff, patients or participants themselves) outside the research group so cannot be used to identify individuals.
Yes
I understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).
Yes
I have followed all appropriate research reporting guidelines and uploaded the relevant EQUATOR Network research reporting checklist(s) and other pertinent material as supplementary files, if applicable.
Yes
Data Availability
Individual participant data collected during the study will be available, after deidentification of all participants. Data will be available to researchers who provide a methodologically sound proposal to achieve the aims in the approved proposal. Proposals should be directed to the corresponding author to gain access to the data. Data requestors will need to sign a data sharing agreement.