1 Abstract
Background Laboratories worldwide face high demands for molecular testing during the SARS-CoV-2 pandemic that might be further aggravated with the upcoming influenza season in the northern hemisphere. Considering that symptoms of influenza are largely undistinguishable from COVID-19, both SARS-CoV-2 and the Influenza viruses require concurrent testing by RT-PCR in patients presenting with symptoms of respiratory tract infection. In this study, we adapted and evaluated a laboratory developed multiplex RT-PCR assay for simultaneous detection of SARS-CoV-2 (dual-target), Influenza-A and Influenza-B (SC2/InflA/InflB-UCT) on a fully automated high-throughput system (cobas6800).
Methods Analytical performance was assessed by serial dilution of quantified reference material and cell culture stocks in transport medium, including pre-treatment for chemical inactivation. For clinical evaluation, residual portions of 164 predetermined patient samples containing SARS-CoV-2 (n=52), Influenza-A (n=43) or Influenza-B (n=19), as well as a set of negative samples was subjected to the novel multiplex assay.
Results The assay demonstrated analytical performance comparable to currently available commercial tests, with limits of detection of 94.9 cp/ml for SARS-CoV-2, 14.6 cp/ml for Influenza-A and 422.3 cp/ml for Influenza-B. Clinical evaluation showed excellent agreement with the comparator assays (sensitivity 98.1%, 97.7% and 100% for Sars-CoV-2, Influenza-A and -B respectively).
Conclusion The SC2/InflA/InflB-UCT allows for efficient high-throughput testing for all three pathogens and thus provides streamlined diagnostics while conserving resources during the Influenza-season.
Highlights
Simultaneous detection of highly pathogenic respiratory viruses Influenza-A/B and SARS-CoV-2
Including a dual-target assay for SARS-CoV-2 detection
Full automation on the cobas6800 high-throughput platform
Competing Interest Statement
ML received speaker honoraria and related travel expenses from Roche Diagnostics. All other authors declare no conflict of interest.
Funding Statement
None
Author Declarations
I confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.
Yes
The details of the IRB/oversight body that provided approval or exemption for the research described are given below:
This work was conducted in accordance with paragraph 12 of the Hamburg hospital law (paragraph 12 HmbKHG). The use of anonymized samples was approved by the ethics committee, Freie und Hansestadt Hamburg, PV5626.
All necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived.
Yes
I understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).
Yes
I have followed all appropriate research reporting guidelines and uploaded the relevant EQUATOR Network research reporting checklist(s) and other pertinent material as supplementary files, if applicable.
Yes
Data Availability
Data available upon request