Abstract
Purpose To develop a generalized signal model for dual-module velocity-selective arterial spin labeling (dm-VSASL) that can integrate arbitrary saturation and inversion profiles.
Theory and Methods A recently developed mathematical framework for single-module VSASL is extended to address the increased complexity of dm-VSASL and to model the use of realistic velocity-selective profiles in the label-control and vascular crushing modules. Expressions for magnetization difference, arterial delivery functions, labeling efficiency, and cerebral blood flow (CBF) estimation error are presented. Sources of error are examined and timing requirements to minimize quantification errors are derived.
Results For ideal velocity-selective profiles, the predicted signals match those of prior work. With realistic profiles, a CBF-dependent estimation error can occur when velocity-selective inversion (VSI) is used for the labeling modules and velocity-selective saturation (VSS) is used for the vascular crushing module. The error reflects a mismatch between the leading and trailing edges of the delivery function for the second bolus and can be minimized by choosing a nominal labeling cutoff velocity that is lower than the nominal saturation cutoff velocity. In the presence of B0 and B1 inhomogeneities, the labeling efficiency of dual-module VSI is more attenuated than that of dual-module VSS.
Conclusion The proposed signal model will enable researchers to more accurately assess and compare the performance of realistic dm-VSASL implementations and improve the quantification of dm-VSASL CBF measures.
Competing Interest Statement
The authors have declared no competing interest.
Funding Statement
This study did not receive any funding
Author Declarations
I confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.
Yes
I confirm that all necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived, and that any patient/participant/sample identifiers included were not known to anyone (e.g., hospital staff, patients or participants themselves) outside the research group so cannot be used to identify individuals.
Yes
I understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).
Yes
I have followed all appropriate research reporting guidelines, such as any relevant EQUATOR Network research reporting checklist(s) and other pertinent material, if applicable.
Yes
Footnotes
minor revisions and added supplementary information