Abstract
Purpose Single-energy computed tomography (CT) often suffers from poor contrast, yet it remains critical for effec-tive radiotherapy treatment. Modern therapy systems are often equipped with both megavoltage (MV) and kilovoltage (kV) x-ray sources and thus already possess the hardware needed for dual-energy (DE) CT. There exists an unexplored potential for enhanced image contrast using MV-kV DE-CT in radiotherapy contexts.
Approach A toy model comprising a single-line integral through a two-material object was designed for computing basis material signal-to-noise ratio (SNR) using estimation theory. Five dose-matched spectra (three kV, two MV) and three variables were considered: spectral combination, spectral dose allocation, and object material composition. The single-line model was extended to a simulated fan-beam CT acquisition of an anthropomorphic phantom with and without a metal implant. Basis material sinograms were computed and synthesized into virtual monoenergetic images (VMIs). MV-kV and kV-kV VMIs were compared with single-energy images.
Results The 80kV-140kV pair typically yielded the best SNRs, but for bone thicknesses greater than 8 cm, the detunedMV-80kV pair surpassed it. Peak MV-kV SNR was achieved with approximately 90% dose allocated to the MV spectrum. For the CT simulations, MV-kV VMIs yielded a higher contrast-to-noise ratio (CNR) than single-energy CT at specific monoenergies. With the metal implant, MV-kV produced a higher maximum CNR and lower minimum root-mean-square-error than kV-kV.
Conclusions This work quantitatively analyzes MV-kV DE-CT imaging and assesses its potential advantages. This technique may yield improved contrast and accuracy relative to dose-matched single-energy CT or kV-kV DE-CT, depending on object composition.
Competing Interest Statement
All authors have completed the ICMJE uniform disclosure form at www.icmje.org/coi_disclosure.pdf and declare: no support from any organization for the submitted work; Patrick La Rivière has received funding from Accuray Inc., Canon Medical Research USA Inc., and Toshiba Medical Research USA, Inc. and is on the Scientific Advisory Board and holds stock options in Metritrack, Inc.; no other relationships or activities that could appear to have influenced the submitted work.
Funding Statement
Accuray Inc. provided early funding of this work.
Author Declarations
I confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.
Yes
I confirm that all necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived, and that any patient/participant/sample identifiers included were not known to anyone (e.g., hospital staff, patients or participants themselves) outside the research group so cannot be used to identify individuals.
Yes
I understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).
Yes
I have followed all appropriate research reporting guidelines, such as any relevant EQUATOR Network research reporting checklist(s) and other pertinent material, if applicable.
Yes
Data Availability
The code used to generate the data in this manuscript can be freely accessed through GitHub at https://github.com/gjadick/dex-single-ray for the single-line integral toy model and https://github.com/gjadick/dex-ct-sim for the fan-beam CT simulation and basis material decomposition.