SUMMARY
Background COVID-19 has stretched the ability of many institutions to supply needed personal protective equipment, especially N95 respirators. N95 decontamination and reuse programs provide one potential solution to this problem. Unfortunately, a comprehensive evaluation of the effects of decontamination on the integrity of various N95 models using a quantitative fit test (QTFT) approach is lacking.
Aims 1) To investigate the effects of up to eight rounds of vaporized H2O2 (VHP) decontamination on the integrity of N95 respirators currently in use in a hospital setting. 2) To examine if N95 respirators worn by one user can adapt to the face shape of a second user with no compromise of integrity following VHP decontamination.
Methods The PortaCount Pro+ Respirator Fit Tester Model 8038 was used to quantitatively define the integrity, measured by fit, of N95 respirators following decontamination with VHP.
Findings There was an observable downward trend in the integrity of Halyard Fluidshield 46727 N95 respirators throughout eight cycles of decontamination with VHP. The integrity of 3M 1870 N95 respirators was significantly reduced after the respirator was worn, decontaminated with VHP, and then quantitatively fit tested on a second user. Furthermore, we uncovered inconsistencies between qualitative fit test and QTFT results that may have strong implications on the fit testing method used by institutions.
Conclusions Our data revealed variability in the integrity of different N95 models after VHP decontamination and exposed potential limitations of N95 decontamination and reuse programs.
Competing Interest Statement
The authors have declared no competing interest.
Funding Statement
This work was supported by University Hospital, Newark NJ; and the National Institute of Allergy and Infectious Diseases of the National Institutes of Health [grant number T32AI125185].
Author Declarations
I confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.
Yes
The details of the IRB/oversight body that provided approval or exemption for the research described are given below:
Experiments involving fit testing and decontamination of N95 respirators was part of the public health response to the COVID-19 pandemic and were thus considered exempt from institutional review board approval. All participants gave informed consent prior to participating in any experiments.
All necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived.
Yes
I understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).
Yes
I have followed all appropriate research reporting guidelines and uploaded the relevant EQUATOR Network research reporting checklist(s) and other pertinent material as supplementary files, if applicable.
Yes
Paper in collection COVID-19 SARS-CoV-2 preprints from medRxiv and bioRxiv
The Chan Zuckerberg Initiative, Cold Spring Harbor Laboratory, the Sergey Brin Family Foundation, California Institute of Technology, Centre National de la Recherche Scientifique, Fred Hutchinson Cancer Center, Imperial College London, Massachusetts Institute of Technology, Stanford University, University of Washington, and Vrije Universiteit Amsterdam.