Abstract
Background Severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) emerges in Wuhan City, Hubei Province, spreads worldwide, and threats the human life. The detection of SARS-CoV-2 is important for the prevention of the outbreak and management of patients. Real-time reverse-transcription polymerase chain reaction (RT-PCR) assay detected the virus in clinical laboratory.
Methods This study utilized primers and single-quencher probes in accordance with the Centers for Disease Control and Prevention (CDC) in the USA and the National Institute of Infectious Diseases (NIID) in Japan. Moreover, we designed the double-quencher probes (YCH assay) according to the oligonucleotide sequence established by NIID. Using these assays, we conducted a one-step real-time RT-PCR with serial DNA positive control to assess the detection sensitivity.
Results The threshold cycle (Ct) value of RT-PCR was relatively low in CDC and YCH assays compared to NIID assay. Serial dilution assay showed that both CDC and YCH assays could detect a low-copy number of DNA positive control. The background fluorescent signal at the baseline was lower in YCH than that of NIID.
Conclusion Double-quencher probes decreased background fluorescent signal and improved detection sensitivity of SARS-CoV-2.
Competing Interest Statement
The authors have declared no competing interest.
Funding Statement
This study was supported by a Grant-in-Aid for the Genome Research Project from Yamanashi Prefecture (to M.O. and Y.H.), the Japan Society for the Promotion of Science (JSPS) KAKENHI Early-Career Scientists (grant number JP18K16292 to Y.H.), a Research Grant for Young Scholars (to Y.H.), the YASUDA Medical Foundation (to Y.H.), the Uehara Memorial Foundation (to Y.H.), and Medical Research Grants from the Takeda Science Foundation (to Y.H.).
Author Declarations
All relevant ethical guidelines have been followed; any necessary IRB and/or ethics committee approvals have been obtained and details of the IRB/oversight body are included in the manuscript.
Yes
All necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived.
Yes
I understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).
Yes
I have followed all appropriate research reporting guidelines and uploaded the relevant EQUATOR Network research reporting checklist(s) and other pertinent material as supplementary files, if applicable.
Yes
Data Availability
Data is available upon request.